Chapter 7 Periodic Maintenance
7-9
7.2.5 Cleaning Brush and Check for Fallen Foreign Objects
Clean the brush (cleaning unit, eraser unit, and toppling prevention block) and the magnet shaft.
Also, confirm if any foreign object has fallen into the equipment.
1
Open the second front door.
2
Remove dust on the brush in the cleaning unit
using a vacuum cleaner.
3
Remove dust on the toppling prevention block
using a vacuum cleaner.
4
Remove dust on the dust removal brush in the
eraser unit using a vacuum cleaner.
1. See the section "6.2.5 Replacing the Erase Lamp
(Halogen Lamp)" in the Operation Manual" to remove
the erase lamp unit.
2. Remove dust on the dust removal brush using a
vacuum cleaner.
3. Attach the eraser lamp unit.
5
Confirm there is no foreign object in the
equipment.
Remove it if there is any. If possible, give client an
advice to be alert.
6
Clean the magnet shaft of the subscan unit.
Wipe off the grim from the magnet shaft using a rag
dampened with a water and squeezed tightly.
This completes the cleaning of the brush and confirmation of the foreign objects.
Go to
"7.2.6 Cleaning Cassette Absorption Magnet (Page 7-10)"
and clean the surface of the magnet
surface of the cassette absorption plate.
Caution
Be sure to turn OFF the power supply circuit breaker and unplug the power cable from the
equipment whenever the maintenance is performed.
Caution
Use the glove while working so there will be
no oil from hand contaminating the filters on
the eraser lamp unit.
Содержание 0902
Страница 2: ......
Страница 6: ......
Страница 12: ......
Страница 30: ...2 10...
Страница 54: ......
Страница 84: ...Chapter 3 Troubleshooting 3 30 3 2 19 Confirm the H sync Signal To be stated...
Страница 85: ...Chapter 3 Troubleshooting 3 31 3 2 20 Confirm the LMC Operation To be stated...
Страница 120: ......
Страница 296: ...8 10...
Страница 300: ......
Страница 301: ......