background image

nm

1310

HZ

270

Auto

-off

nm

1310

HZ

270

Auto

-off

nm

1310

HZ

270

Auto

-off

nm

1310

HZ

270

Auto

-off

KLS-35m stable light source instruction manual

1

3

9

8

4

5

6

12

11

10

7

2

1

3

4

5

2

λ

1310

nm

1550

nm

Auto

-off

HZ

270

HZ

1000

HZ

2000

nm

1310

HZ

270

Auto

-off

CW

CW

External and key function description

OLS interface: laser light source output port

1

LED lighting: flashlight illumination 

2

Display: display test results and other information 

3

Enter key: switch between 7 different output wavelength tests. 

4

LED key: turn on/off the flashlight, press any key to backlight

5

       key: switchable carrier frequency 270Hz/ 1kHz/ 2kHz 

6

On/off key: turn on or off the device 

7

automatic shutdown 

9

frequency 

10

power supply battery

8

unit

11

wavelength

12

CW

Power-on and power-saving function 

settings

         The equipment will turn on 

after pressing the "     " key. Press 

this button again for more than 

two seconds to shut down.
         This instrument has power 

saving function, normal there is no 

operation for about ten minutes 

after booting, The device will 

automatically shut down. If you need

a screen Shield this function to keep 

the equipment working all the time

Status, just short press "on" when 

booting Key, "Auto-off" at the bottom

left of the meter display. Disappearing 

means that the power saving function 

is cancelled.

Backlight control

               In the power-on state, the 

backlight will automatically turn off 

after a period of time. After turning 

off, press any key to activate the 

backlight. The backlight is used to 

illuminate the instrument display at 

night or in dark places.

Wavelength selection

          According to the needs of the 

project, select the corresponding 

wavelength to match the optical power 

meter. After the meter is turned on, 

press the "output wavelength" key, the 

output wavelength and length will be 

switched to the corresponding wave-

length in turn, and will be displayed on 

the display. The laser light source of this 

series can provide 1310nm and 1550nm 

dual-wavelength laser output according 

to demand.

Carrier output

      In the actual project, it is necessary to 

load the audio carrier in the optical signal 

to identify the optical fiber. The equip-

ment contains three carrier frequencies, 

which are 270Hz, 1KHZ, and 2KHz. After 

pressing the "modulated carrier" key, the 

equipment will load different carriers in 

sequence, and the display will be 

displayed at the same time. The current 

carrier frequency will be displayed on the 

screen.

Отзывы: