6 SERVICE OF THE DEVICE STARDEVICE
®
6.1
CLEANING OF THE DEVICE
Regular cleaning of the device ensures reliable and trouble-free operation of the device
STARDEVICE
®
.
You should disconnect the manipulator from the electronic unit and the electronic unit itself from the
mains before beginning any cleaning and / or repair operations.
General external cleaning is performed depending on the frequency of usage and application of the device.
All the details that are in contact with the patient must be treated with medical alcohol wipes or cotton
wool soaked in 70-96% aqueous-alcohol solution.
It is very important to avoid contact with the liquid inside the electronic unit and the manipulator.
It is necessary to keep clean the ventilation slots of the electronic unit and the body of the manipulator.
Only special non-woven napkins for LCD monitors are allowed to clean the TFT display. These wipes
should not contain alcohol.
6.2
CLEANING OF THE TIPS
Cleaning of the tips is performed with the help of medical alcohol wipes or wadded disks, soaked in 70-
96% aqueous-alcoholic solution.
Attention! It is forbidden to clean the tips in any other way that is not described here!
You should clean the tips before each use and after the procedures.
The tips should be stored on a purposefully designed panel.
It is necessary to treat the panel with alcohol solution at least once a month to store the tips.
6.3
REPLACING OF A FUSE
If the device does not work, after connecting to a power source and turn on you should check and replace
a fuse / fuses: maybe one of them is blown. The fuse holders are located on the back panel of the electronic
unit of the device, they have the marking "T4AL250V". Please follow these steps to replace the fuse:
1.
Unplug the power cord from the power socket.
2. Unscrew one of the fuse holders from the panel (marked
"T4AL250V").
3. Remove the blown fuse from the holder.
4. Install a new T4AL250V fuse in the holder.
5. Insert the fuse holder back into the panel and tighten it until it clicks into place.
6. Replace the other fuse by following the instructions above.