KV4000 Kinematic Viscosity Bath 

Operation and Technical Manual 

 

KV4000-Manual 

20 -

 

SHOCK  AND  BURN  HAZARD:

  Only 

clean  inside  the  bath  when  equipment  is  de-
energized and unplugged from the mains power 
supply.  Allow  adequate  time  for  heating  coils  to 
completely cool before cleaning.  

8.3  Draining / Changing Bath Medium 

 

WARNING:

 Do not begin draining the bath 

medium 

until 

it 

has 

reached 

ambient 

temperature.  
 

NOTE:

 When a decision is made to discard this 

material,  dispose  of  this  material  and  its 
container  to  a  hazardous  of  special  waste 
collection  point.  Country,  State,  or  local  laws 
may  impose  additional  regulatory  requirements 
regarding disposal. Dispose in accordance to all 
applicable regulations. 
 

1. 

Remove  one  of  the  seven  port  covers  from 
the top plate of the bath.  

 

2. 

Insert a siphoning tube into the port opening 
and lower tube to bottom of glass bath jar.  

 

3. 

Apply  a  reduced  pressure  to  the  opposite 
end  of  the  tube  with  a  pump  or  withdrawal 
bulb.  

 
NOTE:

  Be  sure  the  discharging  end  of  the 

tube  is  below  the  surface  of  the  bath 
reservoir.  

 

4. 

Remove  the  withdrawal  bulb  or  suction 
pump from the end of the tube and allow the 
bath  medium  to  drain  into  a  lower  reservoir 
or receptacle.  

 

5. 

Once the bath fluid is drained from the bath, 
it  is  time  to  remove  the  complete  top 
assembly of the bath.  

 

6. 

Using  a  flat  head  screwdriver,  remove  the 
four  (4)  silver  screws  located  at  the  four 
corners of the back panel.  

 

7. 

Remove the back panel plate and set aside. 

 

8. 

A collection of wires runs from the lower 
control panel up the left side, while looking 
at the back of the bath. Halfway up the 

wiring are three points where the wires can 
be disconnected. See Figure 4 below:  

 

 

 

Figure 4. Wiring Disconnection Point 

 

9. 

There  are  a  total  of  three  joints  to 
disconnect. First disconnect the single  large 
cluster  of  wires  by  pressing  in  the  sides  of 
the  lower  plastic  piece  then  pulling  the  top 
plastic piece.  

 

10. 

Then disconnect the two smaller joints. Each 
side  of  the  joint  is  labeled  with  an  identical 
number  one  (1)  or  number  two  (2).  This 
makes  for  easy  connection  when  re-
assembling the instrument.  

 

11. 

Once the  wiring  is completely  disassembled 
it  is safe to remove the  top assembly of the 
bath.  

 

12. 

Remove  the  six  screws  lining  the  two  sides 
of  the  top  assembly  with  a  Philips  head 
screw driver. 

 

13. 

Make  sure  there  is  sufficient  space  with 
cleaning  clothes  to  place  the  top  assembly 
when lifted off the bath.  

 

14. 

Lift  the  top  assembly  with  one  hand  on  the 
back side and one hand on the front side of 
the  top  assembly.  Place  top  plate  assembly 
on  flat surface  with the  back  side  of  the  top 
plate  assembly  and  the  heating  coils  on  the 
surface.  

 

15. 

With  the  top  plate  assembly  now  removed 
from  the  main  instrument  it  is  now  easy  to 
access  the  inside  of  the  bath  jar  for  further 
cleaning.  

 

16. 

With the top assembly resting on its back, it 
is  now  possible  to  clean  the  heating  coils, 

Содержание KV4000

Страница 1: ...re Avenue Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Anal...

Страница 2: ......

Страница 3: ...e with the applicable standards set forth in this certification Specifications ASTM D445 ASTM D2532 ASTM D6074 ASTM D6158 IP 71 ISO 3104 DIN 51550 FTM 791 305 NF T 60 100 This unit is tested before it...

Страница 4: ......

Страница 5: ...KV4000 Kinematic Viscosity Bath Operation and Instruction Manual KV4000 Manual 1...

Страница 6: ...KV4000 Kinematic Viscosity Bath Operation and Instruction Manual KV4000 Manual 2...

Страница 7: ...rate statement is given cataloging the items that require separation from the equipment along with basic information on subsequent processing or recycling prior to disposal of the equipment http www d...

Страница 8: ...1716 As a minimum the following substances preparations and components have to be removed from any separately collected WEEE Mercury containing components such as switches or backlighting lamps compac...

Страница 9: ...unning Tests 11 4 3 Temperature Controller Operation 13 5 Operation 14 5 1 Bath 14 5 2 Start Up 15 5 3 Control Panel Operation 15 5 4 Test Procedure 17 6 Safety Features 18 6 1 Over Temperature Protec...

Страница 10: ...cosity Bath Operation and Instruction Manual KV4000 Manual 6 9 Wiring Diagrams 22 9 1 K23702 K23708 115 V Wiring 22 9 2 K23792 K23798 220V Wiring 23 10 Service 24 11 Storage 24 12 Warranty 24 13 Retur...

Страница 11: ...rate and reliable instrumentation as well as support for your laboratory testing programs Please do not hesitate to contact us at any time with your inquiries about equipment tests or technical suppor...

Страница 12: ...for use in accordance with the applicable standard test methods listed in section 1 2 of this manual The use of this equipment in accordance with any other test procedures or for any other purpose is...

Страница 13: ...ment from cart and place on bench 8 Remove foam insert Borosilicate Glass Jar as per steps 9 thru 19 9 Using a flat head screwdriver remove the four 4 silver screws located at the four corners of the...

Страница 14: ...s to these during normal operation can shatter the glass components releasing the bath fluid posing thermal burn hazards and slipping hazards For this reason the instrument must be operated by and acc...

Страница 15: ...y is placed into the bath through the viscometer ports on top of the instrument 7 Bath Bath Stirrer The stirrer constantly circulates bath medium to prevent temperature gradients and ensures temperatu...

Страница 16: ...nt at all temperatures This is advantageous when samples are to be measured at different temperatures Suspended level type viscometers are for transparent liquids of up to 100 000cSt Requires a sample...

Страница 17: ...ase the set point temperature and to increase or change parameter values when programming the temperature controller 6 Exit Function Key This key is used to exit or leave a level One level backward IM...

Страница 18: ...ed for performing kinematic viscosity tests using glass capillary viscometers Please be sure to read the safety and hazard warnings the installation procedure and any of the standard test methods befo...

Страница 19: ...matches that of the ASTM Thermometer Calibrate if necessary See Section 4 3 for full operational details of the temperature controller 5 3 Control Panel Operation Figure 3 Control Panel_Main Screen 1...

Страница 20: ...an be found on the calibration certificate provided with the viscometer 5 Constant 2 Press the Constant 2 button to input a second viscometer tube constant value using the pop up numerical keypad A se...

Страница 21: ...uum system to the initial testing position above the first line This is also called charging the viscometer Place stopper in the top of the viscometer tube and remove the withdrawal bulb NOTE The proc...

Страница 22: ...KV4000 Kinematic Viscosity System is equipped with several safety and protection features which are described in the following sections WARNING The bath is NOT Explosion Proof 6 1 Over Temperature Pro...

Страница 23: ...tenance WARNING Disconnect power to the unit before servicing to avoid exposure to high voltages and or temperatures which may result in personal injury or death If you have any questions about mainta...

Страница 24: ...move the four 4 silver screws located at the four corners of the back panel 7 Remove the back panel plate and set aside 8 A collection of wires runs from the lower control panel up the left side while...

Страница 25: ...lacement Parts Part Number Description 050 001 028 Single Pole Switch 15A 091 032 001 Solid State Relay 4 32 VDC 20A 275 103 036 RTD Temperature Safety Limit 120V 275 103 037 RTD Temperature Safety Li...

Страница 26: ...KV4000 Kinematic Viscosity Bath Operation and Technical Manual KV4000 Manual 22 9 Wiring Diagrams 9 1 K23702 K23708 115 V Wiring...

Страница 27: ...KV4000 Kinematic Viscosity Bath Operation and Technical Manual KV4000 Manual 23 9 2 K23792 K23798 220V Wiring...

Страница 28: ...rn of the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product is the repair or re...

Страница 29: ...KV4000 Kinematic Viscosity Bath Operation and Technical Manual KV4000 Manual 25 Notes...

Страница 30: ...KV4000 Kinematic Viscosity Bath Operation and Technical Manual KV4000 Manual 26 Notes...

Отзывы: