KV4000 Kinematic Viscosity Bath 

Operation and Technical Manual 

 

KV4000-Manual 

10 -

 

18. 

With  the  top  plate  assembly  now  removed 
from  the  main  instrument  it  is  now  easy  to 
access the bath jar to remove the protective 
foam insert.  

 

19. 

Re-assemble the instrument 

 

20. 

Ensure that all parts listed on the packing list 
are  present.  Inspect  the  unit  and  all 
accessories  for  damage.  If  any  damage  is 
found,  keep  all  packing  materials  and 
immediately  report  the  damage  to  the 
carrier. We will assist you with your claim, if 
requested.  When  submitting  a  claim  for 
shipping  damage,  request  that  the  carrier 
inspect 

the 

shipping 

container 

and 

equipment.  Do  not  return  goods  to  Koehler 
without written authorization.  

3.3  Setup 

 

Equipment 

Placement:

 

Make 

sure 

the 

instrument  in  placed  on  a  firm,  level  table  in  an 
area with adequate ventilation or in a hood. The 
unit  may  be  leveled  by  making  minor  turning 
adjustments  to  the  feet  located  at  the  base  of 
the  unit.  Please  note  that  Koehler  does  not 
supply a level with this equipment. 

 

WARNING:

 

Certain  mechanical  stresses 

applied  to  the  tempered  glass  front  panel  and 
inner  the  inner  borosilicate  glass  jar  pose  a 
hazard  to  personnel.  Impacts  to  these  during 
normal 

operation 

can 

shatter 

the 

glass 

components  releasing  the  bath  fluid  posing 
thermal  burn  hazards  and  slipping  hazards.  For 
this  reason  the  instrument  must  be  operated  by 
and  accessible  only  to  trained  and  authorized 
personnel and located in an area where this type 
of impact is likely to occur.  

 
Environmental  Conditions: 

The  instrument 

environment  must  comply  with  the  following 
conditions for proper setup: 

 

  No / Low Dust 

 

  No direct sunlight 

 

  Not near heating or AC ventilation ducts 

 

  No Vibrations 

 

  Clearance from other instruments 

 

  Temperature Range: 5 to 40°C 

 

  Elevation to 2000 meters 

 

  Relative Humidity: < 80% 

 
Ventilation.

  A  fume  hood  or  exhaust  system  is 

required  when  operating  the  unit.  Flammable 
vapors  and/or  steam  are  generated  during 
operation  and  must  not  be  permitted  to 
accumulate. A canopy-style hood may be used if 
the height from the top of the unit to the canopy 
is  5  feet  or  less.  The  exhaust  blower  should 
have a rating of 1000 C.F.M. or greater. 

 
Power: 

Connect the line cords to properly fused 

and  grounded  receptacles  with  the  correct 
voltage  as  indicated  in  section  1.3  or  on  the 
back of the unit.  

 

WARNING:

 

For  safety,  disconnect  the 

power when performing any maintenance and/or 
cleaning.

 

Do 

NOT

  turn  the  power  on  unless  the 

bath is filled with the proper medium; otherwise, 
damage may  occur to the  unit and the  warranty 
will be void. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание KV4000

Страница 1: ...re Avenue Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Anal...

Страница 2: ......

Страница 3: ...e with the applicable standards set forth in this certification Specifications ASTM D445 ASTM D2532 ASTM D6074 ASTM D6158 IP 71 ISO 3104 DIN 51550 FTM 791 305 NF T 60 100 This unit is tested before it...

Страница 4: ......

Страница 5: ...KV4000 Kinematic Viscosity Bath Operation and Instruction Manual KV4000 Manual 1...

Страница 6: ...KV4000 Kinematic Viscosity Bath Operation and Instruction Manual KV4000 Manual 2...

Страница 7: ...rate statement is given cataloging the items that require separation from the equipment along with basic information on subsequent processing or recycling prior to disposal of the equipment http www d...

Страница 8: ...1716 As a minimum the following substances preparations and components have to be removed from any separately collected WEEE Mercury containing components such as switches or backlighting lamps compac...

Страница 9: ...unning Tests 11 4 3 Temperature Controller Operation 13 5 Operation 14 5 1 Bath 14 5 2 Start Up 15 5 3 Control Panel Operation 15 5 4 Test Procedure 17 6 Safety Features 18 6 1 Over Temperature Protec...

Страница 10: ...cosity Bath Operation and Instruction Manual KV4000 Manual 6 9 Wiring Diagrams 22 9 1 K23702 K23708 115 V Wiring 22 9 2 K23792 K23798 220V Wiring 23 10 Service 24 11 Storage 24 12 Warranty 24 13 Retur...

Страница 11: ...rate and reliable instrumentation as well as support for your laboratory testing programs Please do not hesitate to contact us at any time with your inquiries about equipment tests or technical suppor...

Страница 12: ...for use in accordance with the applicable standard test methods listed in section 1 2 of this manual The use of this equipment in accordance with any other test procedures or for any other purpose is...

Страница 13: ...ment from cart and place on bench 8 Remove foam insert Borosilicate Glass Jar as per steps 9 thru 19 9 Using a flat head screwdriver remove the four 4 silver screws located at the four corners of the...

Страница 14: ...s to these during normal operation can shatter the glass components releasing the bath fluid posing thermal burn hazards and slipping hazards For this reason the instrument must be operated by and acc...

Страница 15: ...y is placed into the bath through the viscometer ports on top of the instrument 7 Bath Bath Stirrer The stirrer constantly circulates bath medium to prevent temperature gradients and ensures temperatu...

Страница 16: ...nt at all temperatures This is advantageous when samples are to be measured at different temperatures Suspended level type viscometers are for transparent liquids of up to 100 000cSt Requires a sample...

Страница 17: ...ase the set point temperature and to increase or change parameter values when programming the temperature controller 6 Exit Function Key This key is used to exit or leave a level One level backward IM...

Страница 18: ...ed for performing kinematic viscosity tests using glass capillary viscometers Please be sure to read the safety and hazard warnings the installation procedure and any of the standard test methods befo...

Страница 19: ...matches that of the ASTM Thermometer Calibrate if necessary See Section 4 3 for full operational details of the temperature controller 5 3 Control Panel Operation Figure 3 Control Panel_Main Screen 1...

Страница 20: ...an be found on the calibration certificate provided with the viscometer 5 Constant 2 Press the Constant 2 button to input a second viscometer tube constant value using the pop up numerical keypad A se...

Страница 21: ...uum system to the initial testing position above the first line This is also called charging the viscometer Place stopper in the top of the viscometer tube and remove the withdrawal bulb NOTE The proc...

Страница 22: ...KV4000 Kinematic Viscosity System is equipped with several safety and protection features which are described in the following sections WARNING The bath is NOT Explosion Proof 6 1 Over Temperature Pro...

Страница 23: ...tenance WARNING Disconnect power to the unit before servicing to avoid exposure to high voltages and or temperatures which may result in personal injury or death If you have any questions about mainta...

Страница 24: ...move the four 4 silver screws located at the four corners of the back panel 7 Remove the back panel plate and set aside 8 A collection of wires runs from the lower control panel up the left side while...

Страница 25: ...lacement Parts Part Number Description 050 001 028 Single Pole Switch 15A 091 032 001 Solid State Relay 4 32 VDC 20A 275 103 036 RTD Temperature Safety Limit 120V 275 103 037 RTD Temperature Safety Li...

Страница 26: ...KV4000 Kinematic Viscosity Bath Operation and Technical Manual KV4000 Manual 22 9 Wiring Diagrams 9 1 K23702 K23708 115 V Wiring...

Страница 27: ...KV4000 Kinematic Viscosity Bath Operation and Technical Manual KV4000 Manual 23 9 2 K23792 K23798 220V Wiring...

Страница 28: ...rn of the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product is the repair or re...

Страница 29: ...KV4000 Kinematic Viscosity Bath Operation and Technical Manual KV4000 Manual 25 Notes...

Страница 30: ...KV4000 Kinematic Viscosity Bath Operation and Technical Manual KV4000 Manual 26 Notes...

Отзывы: