KV4000 Kinematic Viscosity Bath
Operation and Technical Manual
KV4000-Manual
-
10 -
18.
With the top plate assembly now removed
from the main instrument it is now easy to
access the bath jar to remove the protective
foam insert.
19.
Re-assemble the instrument
20.
Ensure that all parts listed on the packing list
are present. Inspect the unit and all
accessories for damage. If any damage is
found, keep all packing materials and
immediately report the damage to the
carrier. We will assist you with your claim, if
requested. When submitting a claim for
shipping damage, request that the carrier
inspect
the
shipping
container
and
equipment. Do not return goods to Koehler
without written authorization.
3.3 Setup
Equipment
Placement:
Make
sure
the
instrument in placed on a firm, level table in an
area with adequate ventilation or in a hood. The
unit may be leveled by making minor turning
adjustments to the feet located at the base of
the unit. Please note that Koehler does not
supply a level with this equipment.
WARNING:
Certain mechanical stresses
applied to the tempered glass front panel and
inner the inner borosilicate glass jar pose a
hazard to personnel. Impacts to these during
normal
operation
can
shatter
the
glass
components releasing the bath fluid posing
thermal burn hazards and slipping hazards. For
this reason the instrument must be operated by
and accessible only to trained and authorized
personnel and located in an area where this type
of impact is likely to occur.
Environmental Conditions:
The instrument
environment must comply with the following
conditions for proper setup:
No / Low Dust
No direct sunlight
Not near heating or AC ventilation ducts
No Vibrations
Clearance from other instruments
Temperature Range: 5 to 40°C
Elevation to 2000 meters
Relative Humidity: < 80%
Ventilation.
A fume hood or exhaust system is
required when operating the unit. Flammable
vapors and/or steam are generated during
operation and must not be permitted to
accumulate. A canopy-style hood may be used if
the height from the top of the unit to the canopy
is 5 feet or less. The exhaust blower should
have a rating of 1000 C.F.M. or greater.
Power:
Connect the line cords to properly fused
and grounded receptacles with the correct
voltage as indicated in section 1.3 or on the
back of the unit.
WARNING:
For safety, disconnect the
power when performing any maintenance and/or
cleaning.
Do
NOT
turn the power on unless the
bath is filled with the proper medium; otherwise,
damage may occur to the unit and the warranty
will be void.
Содержание KV4000
Страница 2: ......
Страница 4: ......
Страница 5: ...KV4000 Kinematic Viscosity Bath Operation and Instruction Manual KV4000 Manual 1...
Страница 6: ...KV4000 Kinematic Viscosity Bath Operation and Instruction Manual KV4000 Manual 2...
Страница 27: ...KV4000 Kinematic Viscosity Bath Operation and Technical Manual KV4000 Manual 23 9 2 K23792 K23798 220V Wiring...
Страница 29: ...KV4000 Kinematic Viscosity Bath Operation and Technical Manual KV4000 Manual 25 Notes...
Страница 30: ...KV4000 Kinematic Viscosity Bath Operation and Technical Manual KV4000 Manual 26 Notes...