K47001 Autoignition Apparatus 

Operation and Instruction Manual 

 

K47001-Manual

    

 

-

2-

 

 

1  Introduction 

 

The  Koehler  K47001  Autoignition  Apparatus  is 

used  to  determine  the  hot  and  cool  flame 
autoignition  temperatures  of  a  liquid  in  air  at 
atmospheric  pressure  in  a  uniformly  heated 
vessel. The temperatu

res at which ‘cool flame’ and 

‘hot  flame’  ignitions  occur,  as  evidenced  by 
sudden temperature increases in the sample flask, 
are  measured  and  recorded,  and  the  delay  time 
between introduction of the sample and  ignition  is 
timed.  

 
This  manual  provides  important  information 

regarding  safety,  technical  reference,  installation 
requirements,  operating  condition  specifications, 
user  facility  resource  requirements,  and  operating 
instructions 

for 

the 

Koehler 

Autoignition 

Apparatus.  This  manual  should  also  be  used  in 
conjunction  with  applicable  published  laboratory 
procedures.  Information  on  these  procedures  is 
given in section 1.2. 

 

1.1 

Koehler’s 

Commitment 

to 

Our 

Customers 

 

Providing  quality  testing  instrumentation  and 

technical support services for research and testing 
laboratories  has  been  our  specialty  for  more  than 
50  years.  At  Koehler,  the  primary  focus  of  our 
business  is  providing  you  with  the  full  support  of 
your  laboratory  testing  needs.  Our  products  are 
backed  by  our  staff  of  technically  knowledgeable, 
trained  specialists  who  are  experienced  in  both 
petroleum products testing and instrument service 
to  better  understand  your  requirements  and 
provide  you  with  the  best  solutions.  You  can 
depend on Koehler for a full range of accurate and 
reliable instrumentation as well as support for your 
laboratory  testing  programs.  Please  do  not 
hesitate  to  contact  us  at  any  time  with  your 
inquiries  about  equipment,  tests,  or  technical 
support. 

 

Toll Free: 1-800-878-9070 (US only) 
Tel: +1 631 589 3800 
Fax: +1 631 589 3815 
Email: [email protected] 
http://www.koehlerinstrument.com 

 

 
 
 

1.2  Recommended 

Resources 

and 

Publications 

 

1. 

American  Society  for  Testing  and  Materials 
(ASTM) 
100 Barr Harbor Drive 
West  Conshohocken,  Pennsylvania  19428-
2959, USA 
Tel: +1 610 832 9500  
Fax: +1 610 832 9555 

http://www.astm.org

 

email: [email protected] 

 

ASTM Publication: 

 

ASTM  E659:  Standard  Test  Method  for 
Autoignition 

Temperature 

of 

Liquid 

Chemicals 

 

1.3  Instrument Specifications 

 

Models: 

K47001 

Electrical 

Requirements: 

 
230V 50/60Hz single 
phase 

Temperature Range: 

Ambient to 1200°C 

Temperature Control 

Stability: 

 
± 3 °C 

Open Volume: 

0.16 ft

3

. model: 5 L 

Power: 

2.4 kW 
 

 
 
 
 
 
 
 

Содержание K47001

Страница 1: ...ore Avenue Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Ana...

Страница 2: ...ormance with the applicable standards set forth in this certification Specifications ASTM E659 This unit is tested before it leaves the factory to ensure total functionality and compliance to the abov...

Страница 3: ...rface 5 5 Operation 7 5 1 General Notes 7 5 2 Maintenance and fault clearance 7 5 3 Display Menu Navigation 8 Notifications 19 Control Loops Single Set point Operation 20 Event Control 23 Process Moni...

Страница 4: ...years At Koehler the primary focus of our business is providing you with the full support of your laboratory testing needs Our products are backed by our staff of technically knowledgeable trained sp...

Страница 5: ...at http siri uvm edu or http www sigma aldrich com Disposal Upon delivery the furnace does not contain any material which is to be classified as hazardous waste However process residues may collect in...

Страница 6: ...WARNING For safety disconnect the power when performing any maintenance and or cleaning 3 2 3 Initial Heating Procedure To dry out the brick lining and to create an oxide protection layer on the resi...

Страница 7: ...information help icon will display text based help associated with the current screen Help is available in 28 languages based on the user selection in the offline setup of section of the display The...

Страница 8: ...he next available loop or chart Pressing and holding the arrow keys will continue the item scrolling until the last step of the program is reached or the button is released IMPORTANT Do not use any sh...

Страница 9: ...nd is switched on automatically when the door is closed Do not open the furnace when it is hot If the furnace must be opened at high temperatures do this as quickly as possible Wear personal protectiv...

Страница 10: ...xt Based Navigation Icon Based Slide Page Navigation The available menu items are dependent upon the OEM configuration of the display Not all of the menu items shown may be available on your system Th...

Страница 11: ...old and Halt menu items allow quick access to these functions if a program is already downloaded to the display loop controls The Device menu provides access to the display controller settings These i...

Страница 12: ...splay Settings and Data file functions which include file utilities FTP WAN back up settings as well as the historical data viewer Slide page 2 provides access to automatic ramp soak program actions E...

Страница 13: ...ext Based Program Menu The View menu provides navigation to the program view screens These include the Status Entry and Settings screens The File menu provides functions needed to create open save and...

Страница 14: ...ions needed to create open save and delete automatic ramp soak programs Slide page 2 provides all of the functions needed to edit programs These include Copy Paste Insert and Remove step as well as th...

Страница 15: ...oop Tuning screen The Set menu provides access to settings Alarm set points control loop set point Limits Event tagnames Navigation type selection and Communications options The Email menu provides se...

Страница 16: ...ide page provides access to the display controller settings These include options for Navigation type control loop set point Limits Alarm set points loop Tuning Event tagname entry and Email server se...

Страница 17: ...d ramp soak program is not running and data logging is turned off Text Based Offline Setup Menu The View menu provides navigation back to the display device settings menu The Set menu provides access...

Страница 18: ...Slide page 1 provides navigation back to the device Settings menu Clock Display settings About the display and Exit application screens temperature Units selection Language import export Configuration...

Страница 19: ...n and digital Signature screens The File menu provides functions for opening and deleting historical data log files The Data menu provides access to the main Data screen for starting and stopping data...

Страница 20: ...user can select which items are to be logged Historical data file functions are provided for Opening and Deleting historical files and viewing setting data file Annotations and digital Signatures The...

Страница 21: ...e alarm status including the most recent alarm Data logging status including active file name and the length and logging rate set Security status including the current user Audit trail active inactive...

Страница 22: ...buttons will not be shown The loop name shown at the top right of the loop view will update as each loop is shown The unit display field will update as well in order to provide the user with the unit...

Страница 23: ...ue for cooling Start This button starts the automatic ramp soak program currently loaded in the loop control During program operation the button text will change to Hold Pressing the button will place...

Страница 24: ...g is set to All the current program loaded field for the primary loop will display the program information All other slave loops will indicate that they are under set point control of the primary loop...

Страница 25: ...r contact your OEM for further information Only your OEM can address equipment related issues The names of events are also determined by the system designer The screen shot below is an example showing...

Страница 26: ...rom a minimum of 6 seconds for time periods of 72 minutes or less up to a maximum of 120 seconds for 1440 minutes 24 hours The Labels button allows the user to cycle through each of the configured plo...

Страница 27: ...Right axis The left axis is the default axis selection If none of the channels are assigned to the right axis the axis will not be shown on the chart Once the channel selections have been made press...

Страница 28: ...figuration and operation of your the display controller contact your OEM for further information Only your OEM can address equipment related issues Alarm Monitor The Alarm screen is accessed from the...

Страница 29: ...en to a file Alarm file names are listed as month_day_year The Open file button will display a dialog box where the user can select any alarm file to view NOTE If no alarms occurred on a given day an...

Страница 30: ...ns for creating editing programs Settings Set program operating mode and primary loop File menu New Clears all current program entries Open Provides file open dialog to select a program from memory Sa...

Страница 31: ...nto a master slave program operation While in single set point static mode each loop control operates independently from the others However when an automatic program is started the loop defined as the...

Страница 32: ...racters long However the display only uses the first 14 characters of the program name for display updates and record keeping When naming programs try to keep the name limited to 14 characters in orde...

Страница 33: ...creen Begin by selecting the desired ramp or soak step type available selections are ramp soak jump and end The jump and end step types are discussed in the following sections To enter the set point f...

Страница 34: ...to determine what additional time may be needed in order to make sure that the process reaches set point before starting the soak time Setting Step Set Point and Ramp Rate If the ramp units are set f...

Страница 35: ...ep state and all event outputs will turn off and remain off until the user halts program operation Upon stopping the program the event outputs will then return to their static settings NOTE The step e...

Страница 36: ...made which signifies infinite looping If the user enters 10 000 for the cycle count the program will repeat indefinitely until stopped by the user If the display is configured for automatic program fi...

Страница 37: ...Deviation band hold back is a combination of both the deviation low and high hold back types It places the program in hold when the process value deviates above or below the set point by more then th...

Страница 38: ...revert back to the single set point static mode values that were entered prior to the program being started NOTE When the program operating mode is set to Single and more than one loop is configured...

Страница 39: ...indicates the programmed set point for the loop that will be achieved by the end of the step The field will indicate the current ramp or soak status for the loop based on the set point of the step If...

Страница 40: ...ogram i e present in memory and ready to be run 3 Why is the program not running to the set point entered for the step Automatic ramp soak program operation is limited by the lower and upper set point...

Страница 41: ...residues Insert a new heated chamber insulation element package from the rear of the furnace Open the front door and install the two screws on each side of the vertical stainless steel shelf lip Re co...

Страница 42: ...the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product is the repair or replace...

Страница 43: ...K47001 Autoignition Apparatus Operation and Instruction Manual K47001 Manual 41 Notes...

Страница 44: ...K47001 Autoignition Apparatus Operation and Instruction Manual K47001 Manual 42 Notes...