Operation and Instruction Manual 

 

REV B  

 
 
 
 
 

Koehler Instrument Company, Inc. 

1595 Sycamore Avenue • Bohemia, New York 11716-1796 • USA 

Toll Free: 1-800-878-9070 

(US only) • Tel: +1 631 589 3800 • Fax: +1 631 589 3815 

http://www.koehlerinstrument.com • e-mail: [email protected] 

Petroleum Testing & Analysis Instrumentation • Custom Design & Manufacturing 

 
 
 
 

 

 

 

K3919X 

 Demulsibility Bath 

 

 

 

Содержание K3919 Series

Страница 1: ...camore Avenue Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis Instrumentation Custom Design Manufacturing K3919X Demulsibility Bath ...

Страница 2: ...K3919X Demulsibility Characteristics Bath Operation and Instruction Manual K3919X Manual 1 ...

Страница 3: ...Demulsibility Bath was manufactured in conformance with the applicable standards set forth in this certification Specifications ASTM D2711 This unit is tested before it leaves the factory to ensure total functionality and compliance to the above specifications and ASTM standards Test and inspection records are on file for verification Jesse Kelly Application Engineer Koehler Instrument Company ...

Страница 4: ...K3919X Demulsibility Characteristics Bath Operation and Instruction Manual K3919X Manual 3 ...

Страница 5: ...ller Operation 9 5 Operation 10 5 1 Bath 10 5 2 Power 10 5 3 Install the K39103 Demulsibility Stirrer 11 5 4 Operation Procedure 11 5 5 Cleaning the Stirrer 11 6 Safety Features 11 6 1 Over Temperature Protection 11 6 2 Over Power Protection 11 7 Maintenance 12 7 1 Routine Maintenance 12 7 2 Replacement Parts 13 8 Wiring Diagrams 14 8 1 K39190 Demulsibility Bath 115V 14 8 2 K39199 Demulsibility Ba...

Страница 6: ...K3919X Demulsibility Characteristics Bath Operation and Instruction Manual K3919X Manual 5 12 Warranty 16 13 Returned Goods Policy 17 Notes 18 ...

Страница 7: ...to contact us at any time with your inquiries about equipment tests or technical support Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 Email info koehlerinstrument com http www koehlerinstrument com 1 2 Recommended Resources and Publications 1 American Society for Testing and Materials ASTM 100 Barr Harbor Drive West Conshohocken Pennsylvania 19428 2959 USA Tel 1 610 832 9...

Страница 8: ... 09F 250 000 09C ASTM 9F Thermometer ASTM 9C Thermometer K39170 K39179 Conditioning Bath 115V 60Hz 220 240V 50 60Hz 3 2 Unpacking Carefully unpack and place the instrument and accessories in a secure location Use extra care while unpacking the Pyrex glass jar Ensure that all parts listed on the packing list are present Inspect the unit and all accessories for damage If any damage is found keep all...

Страница 9: ... and left stirrer motors can be changed using the up and down arrows on the tachometer Some fluctuation of the reading is to be expected as the motor turns but the stirrer will meet and exceed the ASTM specifications At the end of the run the tachometer may continue to display a reading for about 30 seconds Please ignore this as the Tachometer will reset to zero momentarily 6 Stirrer The Demulsibi...

Страница 10: ...unnels in two removable 4 unit racks for conditioning prior to testing in Demulsibility Apparatus Includes microprocessor digital temperature control automatic water level control and gabled cover 360 000 003 Digital Tachometer Handheld non contact LCD tachometer takes measurements up to 3 ft away with 1rpm accuracy Supplied with four 1 5V AA batteries 250 000 09F 250 000 09C ASTM 9F 9C Thermomete...

Страница 11: ...shing If further adjustments are necessary press the Program Key again Resume regular operations by pressing the Exit Function key two times Verify if the new calibration is correct by observing the upper red display and comparing the value with the calibrated reference device Auto Tune This routine allows the digital temperature control to learn the heating parameters needed for any particular se...

Страница 12: ...An instruction manual is provided with the tachometer 5 Proceed to test in accordance with an applicable standard test method 5 5 Cleaning the Stirrer When the test procedure has been completed clean the stirrer by immersing it in the solvent tank containing 1 1 1 Trichloroethane and operating it at high speed for several minutes WARNING The use of flammable cleaning solvents is extremely hazardou...

Страница 13: ... your equipment then please do not hesitate to contact the Koehler technical service department 7 1 Routine Maintenance The K3919X Demulsibility Apparatus requires little routine maintenance to provide many years of continuous service However over the course of time some instrument parts may need to be replaced When ordering replacement part s please provide the model number serial number and prod...

Страница 14: ...emp Probe x 12 381 115 003 Electronic Timer 115V 50 60 Hz 381 230 003 Electronic Timer 230V 50 60 Hz 091 032 001 Solid State Relay 4 31V DC 20A K23700 03013A Motor 115V 60Hz 1 20HP SS Shaft K23700 03014A Motor 230V 50 60HZ 1 15HP K39180 0 10 Heater 1000W 115V K39189 0 10 Heater 1000W 230V 360 115 016 Digital Tachometer 275 020 001 Motor Controller 090 120 014 Relay 120V 10A SPDT 090 240 018 Relay ...

Страница 15: ...K3919X Demulsibility Characteristics Bath Operation and Instruction Manual K3919X Manual 14 8 Wiring Diagrams 8 1 K39190 Demulsibility Bath 115V ...

Страница 16: ...K3919X Demulsibility Characteristics Bath Operation and Instruction Manual K3919X Manual 15 8 2 K39199 Demulsibility Bath 220 240V ...

Страница 17: ...the K3919X Demulsibility Apparatus should not require service Any service problem can be quickly resolved by contacting Koehler s technical service department either by letter phone fax or email In order to assure the fastest possible service please provide us with the following information Model Number Serial Number Date of Shipment 11 Storage This laboratory test instrument is equipped with elec...

Страница 18: ...e item s to be returned and the reason for the return You will be issued a Returned Authorization RA number which must be prominently displayed on the shipping container when you return the material to our plant Shipping containers without an RA number prominently displayed with will be returned to the sender Goods must be returned freight prepaid Returns will be subject to a restocking charge the...

Страница 19: ...K3919X Demulsibility Characteristics Bath Operation and Instruction Manual K3919X Manual 18 Notes ...

Страница 20: ...K3919X Demulsibility Characteristics Bath Operation and Instruction Manual K3919X Manual 19 Notes ...