K2269X Low Temperature Pressure Viscometer 

Operation and Instruction Manual 

 

K2269X-Manual

    

 

-

7-

 

16.  Chamber  Temperature  Controller.

  The 

temperature  controller  regulates  the  chamber 
temperature  for  the  test  procedure.  Set  point 
temperature 

is 

in 

green 

and 

actual 

temperature  of  the  chamber  is  displayed  in 
red.  Refer  to  Section  4.5  for  full  operational 
details. 

 

17.  Sample Temperature Display.

 For displaying 

the  sample  temperature  of  the  grease  during 
testing. 

 

 

4.4  Gear Box and Motor Descriptions 

 

 

 

Figure 5. Instrument Descriptions_Gear Box and Motor 

 

18.  Gear  Box  Cover.

  Protects  user  from  rotating 

gears.  Lift  gear  box  to  access  gears  for 
removal of ice / frost build-up. When gear box 
cover  is  lifted,  drive  motor  will  automatically 
stop and gears will stop rotating. Upon closing 
gear box cover, drive motor and gears should 
start up automatically. In the event the system 
does  not  restart,  press  the  red  Reset  button 
located on the control box, See Figure 4, Item 
14.  

 

19.  Speed Reducer:

 200 to 1 ratio speed reducer 

is coupled to the drive motor to obtain desired 
speed  for  gears  used  to  power  the  hydraulic 
pump system.  

 

20.  Variable  Speed  Drive  Motor: 

1/8-hp,  3000 

rpm  motor  is  powered  on  by  pressing  the 
Drive  Motor  /  Circulating  Pump  switch  to  the 

ON

  position.  See  Figure  4,  item  13.  Variable 

Speed  Motor  is  coupled  to  the  speed  reducer 
and  is  capable  of  operating  at  two  speeds 

required to simulate flow rates attained by 40-
tooth  gear  and  64-tooth  gear  as  per  ASTM 
D1092.  Therefore,  it  is  not  required  to 
physically  switch  gear  from  40-tooth  to  64-
tooth type and vice-versa. 

 

 
21.  Refrigerant  Circulation  Piping. 

Circulating 

Pump  located  in  the  Refrigerated  Test 
Chamber  pushes  refrigerant  through  the 
refrigerant  circulating  piping,  cooling  the 
hydraulic fluid and gear pump, and flows back 
into the Refrigerated Test Chamber.  

 
22.  Over-Pressure  Valve. 

In  the  event,  the 

pressure of the hydraulic system becomes too 
high,  the  over-pressure  valve  will  burst  and 
safely release the pressure inside the system. 

 
23.  Gear  Pump. 

Pushes  hydraulic  fluid  through 

the system. Gear pump is insulated to prevent 
loss  of  cooling  from  the  refrigeration  system. 
The  gear  pump  must  be  as  close  to  thermal 
equilibrium  with  the  refrigeration  system  as 
possible  to  properly  pump  the  hydraulic  fluid 
throughout the system.  

 
24.  Gear. 

Dual  gear  set  is  fixed  to  the  speed 

reducer  and  gear  pump  to  provide  the  proper 
flow  rate  of  the  hydraulic  fluid  through  the 
system. It is not necessary to remove the gear 
at  any  time.  The  two  fluid  flow  rates  required 
by the ASTM test method can be obtained by 
flipping  the  Gear  Switch  on  the  control  box. 
See Figure 4, Item 15.  

 

4.5  Temperature Controller Descriptions 

 

 

 

Figure 4. Temperature Controller 

 

Содержание K2269X

Страница 1: ...enue Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis...

Страница 2: ......

Страница 3: ...tured in conformance with the applicable standards set forth in this certification Specifications ASTM D1092 This unit is tested before it leaves the factory to ensure total functionality and complian...

Страница 4: ......

Страница 5: ...tion 5 4 1 Instrument Descriptions 5 4 2 Top Assembly Description 5 4 3 Control Box Descriptions 6 4 4 Gear Box and Motor Descriptions 7 4 5 Temperature Controller Descriptions 7 5 Operation 9 5 1 Ref...

Страница 6: ...pport of your laboratory testing needs Our products are backed by our staff of technically knowledgeable trained specialists who are experienced in both petroleum products testing and instrument servi...

Страница 7: ...t may exhibit potential hazards Any user must be familiarized with the possible dangers before use We also recommend consulting the Material Data and Safety Sheet MSDS on each chemical reagent for add...

Страница 8: ...e unit and all accessories for damage If any damage is found keep all packing materials and immediately report the damage to the carrier We will assist you with your claim if requested When submitting...

Страница 9: ...ar pump gears drive motor required to power the hydraulic system Refer to section 4 4 for detailed descriptions on this section of the apparatus 4 2 Top Assembly Description Figure 3 Instrument Descri...

Страница 10: ...nect the grease cylinder to the hydraulic system This port is also used as the main connection point for the hydraulic system calibration by utilizing the Calibration Pipe Assembly See Figure 1 Item G...

Страница 11: ...n See Figure 4 item 13 Variable Speed Motor is coupled to the speed reducer and is capable of operating at two speeds required to simulate flow rates attained by 40 tooth gear and 64 tooth gear as per...

Страница 12: ...a certified calibrated measuring device to acquire the temperature Calculate the difference between the measuring device and the Process value displayed on the controller b Press the program key two t...

Страница 13: ...ion instructions for the temperature controller NOTE Allow for the test chamber to cool and come to temperature equilibrium prior to turning on Drive Motor Recirculating Pump See Figure 4 Item 13 It m...

Страница 14: ...imed 11 Once a steady flow or constant drip rate is observed the hydraulic system is primed and free of air bubbles 12 Using the relief valve at the right hand side of the calibration piping assembly...

Страница 15: ...esting purposes If not attached ferrule will be in thermocouple packaging 3 Remove bottom cap from Grease Cylinder F 4 Invert grease cylinder and place Grease Cylinder Piston C inside Grease Cylinder...

Страница 16: ...cylinder loading procedure is now complete Before attaching a capillary changing the flow of the hydraulic fluid and starting a test the sample grease must attain temperature equilibrium with the test...

Страница 17: ...may result in personal injury or death If you have any questions about maintaining your equipment then please do not hesitate to contact the Koehler technical service department 6 1 Routine Maintenan...

Страница 18: ...104 002 Fuse 25A Slo Blo 5X20 M M 220 240 013 Cartridge Heater 500W 240V 265 500 001 RTD Temperature Probe x 12 275 103 044 Temperature Controller 100 240V 288 230 018 Motor 208 230V 3 Phase 288 230 0...

Страница 19: ...K2269X Low Temperature Pressure Viscometer Operation and Instruction Manual K2269X Manual 15 7 Wiring Diagrams 7 1 220 240V 50Hz Unit Wiring K22695...

Страница 20: ...e the return of the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product is the re...

Страница 21: ...K2269X Low Temperature Pressure Viscometer Operation and Instruction Manual K2269X Manual 17 Notes...

Страница 22: ...K2269X Low Temperature Pressure Viscometer Operation and Instruction Manual K2269X Manual 18 Notes...

Отзывы: