Low Temperature Torque Apparatus  

Operation and Instruction Manual 

 

K18854,55,64,65-Manual

    

 

-

12-

 

4  Descriptions

 

 

4.1  Instrument Controls 

 

 

 

 

1.  Power Switch.

 This switch controls the power 

to the entire unit. When the power switch is in 
the 

ON

  position,  the  digital  temperature 

controller,  the  torque  meters,  compressors 
and drive systems of the unit are powered on. 

 

2.  Temperature  Controller.

  The  temperature 

controller  regulates  the  chamber  temperature 
for the test procedure. Refer to Section 4.2 for 
full operational details. 
 

3.  Spindle 

Temperature: 

Displays 

the 

temperature of the interior of the ASTM D4693 
test rig in °C.  

 

4.  Left Drive Motor Switch: 

Controls the ASTM 

D1478 test rig motor drive

 

 

5.  Right  Drive  Motor  Switch.

  Controls  the 

ASTM D4693 test rig motor drive 
 

 

 

6.  Torque: 

Two  meters  display  the 

load

  values 

(NOT  THE  TORQUE  VALUES)  from  the  test 
rigs in 

lbs

, arranged as follows: 

 

ASTM D1478 Test Rig 

– Left side meter 

ASTM D4693 Test Rig 

– Right side meter 

 

Over-torque  cut-off  feature 

–  The  torque 

meter  is  equipped  with  an  automatic  cut-off 
that  interrupts  power  to  the  drive  motor  when 
measured  torque  exceeds  a  programmed 
limit.  This  limit  is  factory  preset  at  55  lbs  to 
protect  the  load  cell  and  drive  system 
components.  
 
To  reset  the  drive  unit  after  cut-off,  depress 
both  buttons  on  the  face  of  the  meter 
simultaneously and release.  
 
To charge the cut-off limit: 
a) Depress  the  button  on  the  right side  of  the 
meter to confirm the current cut-off valve.  
b)  While  depressing  the  button,  adjust  the 
value by turning the recessed set screw on the 
face  of  the  torque  meter  with  a  small  slotted 
screwdriver.  

CAUTION

 

–  Do  not  adjust  the  cut-off  value 

higher  than  55  lbs  or  damage  to  the  drive 
system  and/or  load  cell  may  occur  during 
operation.  

 

4.2  Temperature Controller Operation

 

 

 

 

 
 

Figure 4. Temperature Controller 

 

1.  Process  Temperature  Display.

  The 

upper red LED display shows the process 
temperature as read from the RTD probe. 

 

2.  Set  Point  Temperature  Display.

  The 

lower  green  LED  display  shows  the  set 
point temperature of the controller. 
 

Содержание K18854

Страница 1: ...mia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis Instrumen...

Страница 2: ...Low Temperature Torque Apparatus Operation and Instruction Manual K18854 55 64 65 Manual 1...

Страница 3: ...rque Apparatus was manufactured in conformance with the applicable standards set forth in this certification Specifications ASTM D1478 ASTM D4693 ASTM D4950 FTM 791 334 This unit is tested before it l...

Страница 4: ...Low Temperature Torque Apparatus Operation and Instruction Manual K18854 55 64 65 Manual 3...

Страница 5: ...Low Temperature Torque Apparatus Operation and Instruction Manual K18854 55 64 65 Manual 4...

Страница 6: ...Low Temperature Torque Apparatus Operation and Instruction Manual K18854 55 64 65 Manual 5...

Страница 7: ...information on subsequent processing or recycling prior to disposal of the equipment http www dpa system dk en WEEE Products Private or professional use Responsibility for Registration and Annual Repo...

Страница 8: ...ely collected WEEE Mercury containing components such as switches or backlighting lamps compact fluorescent lamps CFL Batteries Printed circuit boards if the surface of the printed circuit board is gr...

Страница 9: ...1 3 3 Setup 11 4 Descriptions 12 4 1 Instrument Controls 12 4 2 Temperature Controller Operation 12 5 Operation 14 5 1 Preparation for Operation 14 5 2 Operating Instructions 14 5 3 Calibration Proced...

Страница 10: ...g needs Our products are backed by our staff of technically knowledgeable trained specialists who are experienced in both petroleum products testing and instrument service to better understand your re...

Страница 11: ...instructions for preparing the equipment assume that the user is aware of the contents of this document which lists the warranty conditions and important precautions 3 1 Packing List Low Temperature T...

Страница 12: ...medium otherwise damage may occur to the unit and the warranty will be void It is strongly suggested that a voltage stabilizer be used to prevent voltage variability The unstable voltage may cause dam...

Страница 13: ...TM D4693 Test Rig Right side meter Over torque cut off feature The torque meter is equipped with an automatic cut off that interrupts power to the drive motor when measured torque exceeds a programmed...

Страница 14: ...g device and the Process value displayed on the controller b Press the program key two times until PCt is displayed in the lower green LED display Press the DOWN key CAL will display on the lower gree...

Страница 15: ...followed by the outer compression plate with the shoulder facing the hub Slip on the washer and thread on the nut but do not tighten 6 Carefully mount the test assembly in a vertical position in a vi...

Страница 16: ...st Chamber After testing has been completed and the refrigeration system shuts down allow the test chamber to return to room temperature before wiping it dry with a clean cloth This will minimize fros...

Страница 17: ...rod and slip on the weight marked 1 The strain gauge should display 11 lbs Each additional weight should register an additional 11 lbs Adjust the span screw to make any correction NOTE Handle weights...

Страница 18: ...a suitable solvent and air dry Do not permit wet rigs to sit out as the components will corrode Wipe down the exterior of the cabinet using a soft cloth sprayed with household cleaner Over the course...

Страница 19: ...Low Temperature Torque Apparatus Operation and Instruction Manual K18854 55 64 65 Manual 18 8 Wiring Diagrams 8 1 K18864 Unit Wiring 220 240V 60Hz...

Страница 20: ...Low Temperature Torque Apparatus Operation and Instruction Manual K18854 55 64 65 Manual 19 8 2 K18865 Unit Wiring 220 240V 60Hz...

Страница 21: ...turn of the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product is the repair or...

Страница 22: ...Low Temperature Torque Apparatus Operation and Instruction Manual K18854 55 64 65 Manual 21 Notes...

Отзывы: