Section 2: Operating the AUTOMIXER

Low-Level Warning System

The components of the Low-Level Warning System are Solution-Level and Low-Level Indicators and an audible
alarm. The Solution-Level Indicators illuminate to indicate a corresponding volume of solution in the replenishment
tanks. When it is necessary to add more chemicals, the appropriate Low-Level Indicator illuminates and an audible
alarm operates.

Note

• The Low-Level Warning System resets when the mixing starts.

• If the operator is not ready to make a mix and the Low-Level Alarm is operating, open the Drawer until the sound

stops.

• If it is necessary to increase or decrease the sound of the Low-Level Alarm, contact qualified service personnel.

Making a Mix

Caution

• Opening the Developer or Fixer Drawers may cause light leakage into the darkroom in a through-the-wall

installation.

• Prevent damage to the slide mechanism. Do not use excessive force when opening the Drawers.

LOW

LEVEL

FULL

1/2

1/4

FULL

1/2

1/4

LOW

LEVEL

H115_0003HA

DEVELOPER DRAWER

SOLUTION LEVEL

LOW-LEVEL
INDICATOR

DEVELOPER

INDICATORS

H115_0003HCA

INDICATOR

LOW-LEVEL

FIXER

DRAWER

FIXER

Содержание AUTOMIXER III PLUS

Страница 1: ...Eastman Kodak Company Kodak Home Page on Internet Tech Bulletin on Intranet Table of Contents Publication No 9B8815 January 1997 OPERATOR MANUAL for the Kodak AUTOMIXER III PLUS...

Страница 2: ...e this information without notice and makes no warranty express or implied with respect to this information Kodak shall not be liable for any loss or damage including consequential or special damages...

Страница 3: ...interference This digital apparatus does not exceed the Class A limits for radio noise emissions from digital apparatus as set out in the interference causing equipment standard entitled Digital Appar...

Страница 4: ...stem resets when the mixing starts If the operator is not ready to make a mix and the Low Level Alarm is operating open the Drawer until the sound stops If it is necessary to increase or decrease the...

Страница 5: ...ackaged in recyclable plastic Bottles Look for the mark on the bottom of the Bottles for correct disposal The RA developer concentrate made by Kodak makes a 9 5 L 21 2 gal mix To fill the AUTOMIXER to...

Страница 6: ...2 to 3 minutes Weekly Maintenance To keep the AUTOMIXER in optimum operating condition minimum weekly maintenance is required Do the following procedure 1 Open the Drawers Caution Prevent dilution of...

Страница 7: ...does not cover circumstances beyond the control of Kodak misuse abuse attachments accessories alterations not marketed by Kodak including service or parts to correct problems resulting from the use o...

Страница 8: ...OPERATORS MANUAL 8 January 1997 9B8815 This page is intentionally blank...

Страница 9: ...Publication History 9B8815 January 1997 9 Section 4 Publication History Print Date Pub No Affected Pages File Name Notes January 1997 9B8815 All om3402_1 doc First Printing...

Страница 10: ...om3402_1 fm Printed In USA Health Imaging EASTMAN KODAK COMPANY ROCHESTER N Y 14650 Kodak is a trademarks...

Отзывы: