Operating Instructions 

for 

Digital Manometer 

 

Model: MAN-SD 

 

Содержание MAN-SD series

Страница 1: ...Operating Instructions for Digital Manometer Model MAN SD...

Страница 2: ...ectrical Connection 6 8 Function Keys 7 9 Adjustments 7 10 Control Functions 8 11 Maintenance 12 12 Technical Information 12 13 Order Codes 13 14 Dimensions 14 15 Declaration of Conformance 15 Manufac...

Страница 3: ...dents When used in machines the measuring unit should be used only when the machines fulfil the EWG machine guidelines 3 Instrument Inspection Instruments are inspected before shipping and sent out in...

Страница 4: ...nes and systems These devices are equipped as follows 4 digit LCD display Three Programming keys Process connection St Steel Power Supply via 9V block battery Limit relays option Peak value memory opt...

Страница 5: ...that the piping is fully depressurised The digital manometer should be mounted just like a mechanical manometer With standard thread connection sealing is achieved by means of a suitable gasket flat s...

Страница 6: ...the unit and connect the 9V block battery with the connection plug Place the 9V block battery in the enclosure and close it with the lid Terminate the connection wires on the plug cable as shown in t...

Страница 7: ...ds P Enter value and jump to next Menu Option Reject input Return to Menu Option 9 Adjustments Possible adjustments of the device 1 Zero point 2 Password factory pre set 0005 3 Peak value memory optio...

Страница 8: ...hon LCDDisplay 1234 P 1xP Unit Switchoff Passwordinput PASS P 0000 Password set on0005 factorysetting Cancel P AccesstofurtherMenuOptions Automaticzero point setting ZERO P SEt Onconfirmation automati...

Страница 9: ...Peak value memory 1xP Unit switched off Peak value memory P d u P o F F Last Setting m A x Register max value m I n Register min value o F F Peak value memory Off Select Cancel P Confirm selection an...

Страница 10: ...ng pressure r E f P 0 6 5 0 Last switching point setting Set sw point on 07 50 example Cancel P Relay set point on falling pressure r E b P 0 6 4 0 Last return switching point setting Set return sw po...

Страница 11: ...k value memory Off Select Cancel P Confirm selection and go to next Menu Option Password input P A S S P 0 0 0 0 Password Set on 0005 factory setting Cancel P Access to further Menu Options Relay set...

Страница 12: ...o 10 per sec factory adjustable Automatic switch off times 2 90 min only factory adjustable auto off not recommended with analogue or switching output 0 inactive Zero point correction 25 Parts in cont...

Страница 13: ...thout 0 with S Measuring range code 1 to 0 bar A315 0 to 1 bar B025 0 to 2 5 bar B045 0 to 4 bar B055 0 to 6 bar B065 0 to 10 bar B075 0 to 16 bar B085 0 to 25 bar B095 0 to 40 bar B105 0 to 60 bar A1...

Страница 14: ...MAN SD page 14 MAN SD 03 04 14 Dimensions 51 14 74 G1 4A M12 x 1 24 Rotatable Display 3 5 5 Optional PG7 LCD SW 24 89 38 2 30...

Страница 15: ...s with the standards given below EN 50081 1 2 1994 03 Electromagnetic compatibility Generic emission standard EN 61326 1 1998 01 EN 61010 1 1994 03 Safety requirements for electrical measurement contr...