![Kobold Heinrichs Coriolis Скачать руководство пользователя страница 57](http://html1.mh-extra.com/html/kobold/heinrichs-coriolis/heinrichs-coriolis_operating-manual_1982428057.webp)
OPERATING MANUAL HPC & UMC4
Page 57 of 105
14.4.4 Zero point calibration
Using the
Zero point calibration
function the operator can recalibrate the zero point of your meter in the
measuring system. Zero point calibration is to be realized after any installation procedure or after any type
of work has been performed on in the pipes near the sensor.
CAUTION:
This function may only be carried out if it is certain that the fluid in the sensor is not flow-
ing. Otherwise, the flow rates measured subsequently will be incorrect. The sensor may
be completely empty or filled with fluid. A partially filled sensor or air bubbles will lead to
an incorrect zero point calibration.
Calibrating a sensor filled with a fluid is better than calibrating an empty one.
After choosing the
Zero point calibration
function and pressing
„Enter“, the current remaining flow will be
displayed:
QM = 0,00 kg/h
cal. ? [no]
According to the description in Section13.4.3.1 Selection window/make a selection, the operator can tog-
gle between [yes] and [no]. After setting the new value, press
„Enter“ to confirm your entry. Enter [yes] to
have the zero point recalibrated.
14.4.5 Filter
For a noise reduction of the actual readings and minimizing disturbances on the sensor lines signal filter
are used. Following settings are available:
weak
mid
standard (factory setting)
strong
special
Selecting of
“weak”, “mid” or standard influences the dynamics of the actual readings just less. The set-
ting of damping time (see section 14.4.1
“Damping” at page 56 above) determines the dynamic behavior
of sensor and transmitter. Filter “strong” operates as an intense low pass filter to reduce the noise of the
actual readings. Then if setting the damping time shorter than 3 seconds, the low pass filter determines
the dynamic behavior to actua
l reading’s variations.
Note:
Setting of “special 1” to “special 4” is performed by the manufacturer. This setting should
not be changed or selected in standard applications. These filter characteristics are op-
timized for custom specific applications, designed by the manufacturer. Using these fil-
ters in standard applications, the result a degrading of the measurement signal!
According to the description in section 13.4.3.1 Selection window/make a selection, the operator can tog-
gle between noise filter types. After setting the new value, press
“Enter” to confirm your entry.