36
3.
When the Aux 1 purge air
control valve
(C)
is connected,
unscrew the sealing cap and
connect the second transparent
tube here.
Tighten the coupling nut hand
tight.
You can use purge air to
remove residues (process
medium, buffer, cleaning agent)
from the calibration chamber of
the retractable fitting.
8.3 Media Supply to Unical 9000(X)
Connecting the Tubes
C
4.
Connect the green tubes
to the control valve of the
retractable fitting
(A)
.
Left:
“Process“ position
Tube with large diameter
(Ø 8 mm)
Right:
“Service“ position
Tube with small diameter
(Ø 6 mm)
Retractable Fitting (Probe)
Aux 1 Purge Air (Option)