![Knick Stratos Pro A401B PH Скачать руководство пользователя страница 87](http://html.mh-extra.com/html/knick/stratos-pro-a401b-ph/stratos-pro-a401b-ph_user-manual_1980199087.webp)
87
Configuration
Menu item
Action
Choices
Temp compensation,
process medium
For pH measurement
only: Select temperature
compensation of the
process medium.
Linear: LIN
Ultrapure water:
PUREWTR
Table: USERTAB
Select using
key,
proceed using enter.
Temperature
compensation, linear
Enter the linear tempera-
ture compensation of the
process medium.
Enter value using
keys.
Press enter to confirm.
Temperature
compensation, table
Confirm safety
prompt,
then enter values
(5°C step size)
When you have selected
temperature compensa-
tion via table (USERTAB),
you can enter values for
a TC table from 0 to 95 °C
in 5-K steps.
The analyzer displays
temperature values in
5 °C steps. You must spec-
ify the percent deviation
of the measured value
from each of these tem-
perature values.
Intermediate values are
linearly interpolated.
TC compensation can be
defined separately for
parameter sets A and B.
0 ... 95 °C
(5°C step size)
5
Содержание Stratos Pro A401B PH
Страница 1: ...www knick de Latest Product Information Stratos Pro A401B PH User Manual 094698 ...
Страница 11: ...11 Typical Application of Stratos Pro A401B PH IrDA ...
Страница 42: ...42 ...
Страница 50: ...50 Configuration Select Default Sensor SENSOR ISM for ISM sen sors only Configuration ...
Страница 53: ...53 Configuration Configuration Choices Default Relay 1 2 RL1 RL2 Selected in text line ...
Страница 104: ...104 ...
Страница 148: ...148 ...
Страница 167: ...167 Buffer Tables 10 DIN 19267 buffers ...