81
Configuration
Menu item
Action
Choices
Output current for
error message (FAIL)
In the case of an error
(FAIL), the current output is
set to 22 mA.
Select ON or OFF using
keys.
Press enter to confirm.
Output current for
Sensoface (FACE)
In the case of a Sensoface
alert (FACE), the current
output is set to 22 mA.
Select ON or OFF using
keys.
Press enter to confirm.
pH
Temp
Display
Error (22 mA)
Sensoface alert (22 mA)
Error messages and Sensoface alerts can be set separately for both current outputs.
This allows, for example, signaling error messages only over current output 1 and
Sensoface alerts only over current output 2.