![Knick Stratos Pro A401B PH Скачать руководство пользователя страница 115](http://html.mh-extra.com/html/knick/stratos-pro-a401b-ph/stratos-pro-a401b-ph_user-manual_1980199115.webp)
115
Time and Date
Control of the calibration and cleaning cycles is based on the time and
date of the integrated real-time clock.
In measuring mode the time is shown in the lower display.
When using digital sensors, the calibration data is written in the sensor
head.
In addition, the logbook entries (cf Diagnostics) are provided with a
time stamp.
Please note:
• After prolonged power outage (> 5 days) the time display is
replaced by dashes and cannot be used for processing.
Enter the correct time.
• There is no automatic switchover from winter to summer time!
Be sure to manually adjust the time!
Configuration
Menu item
Action
Choices
Tag number
Select character using
keys,
select next digit using
keys.
Press enter to confirm.
A...Z, 0...9, – + < > ? / @
The first 10 characters are
seen in the display with-
out scrolling.
Tag Number (“ TAG“)
You can enter a designation for the point of measurement (tag
number) in the lower display line. Up to 32 digits are possible.
Pressing meas (repeatedly) in the measuring mode indicates the tag
number.
Being part of the device configuration, the “TAG“ can be read out via
IrDA.
A standardized tag number helps, for example, to correctly re-install a
device after repair.
5
Содержание Stratos Pro A401B PH
Страница 1: ...www knick de Latest Product Information Stratos Pro A401B PH User Manual 094698 ...
Страница 11: ...11 Typical Application of Stratos Pro A401B PH IrDA ...
Страница 42: ...42 ...
Страница 50: ...50 Configuration Select Default Sensor SENSOR ISM for ISM sen sors only Configuration ...
Страница 53: ...53 Configuration Configuration Choices Default Relay 1 2 RL1 RL2 Selected in text line ...
Страница 104: ...104 ...
Страница 148: ...148 ...
Страница 167: ...167 Buffer Tables 10 DIN 19267 buffers ...