Calibration with Calibration Solution
Input of temperature-corrected value of calibration solution with
simultaneous display of cell constant
Display
Action
Remark
Select Calibration.
Press
enter
to proceed.
Select CAL_SOL
calibration method.
Press
enter
to proceed.
Ready for calibration.
Hourglass blinks.
Display (3 sec)
Now the device is in
HOLD mode.
Immerse sensor in
calibration solution.
Enter the temperature-
corrected value of the
calibration solution
using the arrow keys
(see table).
Press
enter
to confirm.
Lower line: Display
of cell constant and
temperature
The determined cell
constant is displayed.
The “hourglass” icon is
blinking.
Press
enter
to proceed.