127
Error Handling
Alarm condition:
• The display backlighting turns
red
• The alarm icon
is displayed
• The complete measured-value display blinks
• “
ERR xxx
“ is displayed in the lower menu line
Press the [
info
] key to view a short error text:
• The error text appears in the lower menu line
• The main display reads “
InFo
“.
Parameter errors:
Configuration data such as current range, limit values, etc are checked
during the input.
If they are out of range,
• “
ERR xxx
“ is displayed for 3 sec,
• the display backlighting flashes red,
• the respective maximum or minimum value is shown,
• input must be repeated.
If a faulty parameter arrives through the interface (HART),
• an error message will be displayed: “
ERR 100...199
“
• the faulty parameter can be localized by pressing the [
info
] key
Calibration errors:
If errors occur during calibration,
• an error message will be displayed
Sensoface:
If the Sensoface becomes sad,
• the display backlighting will turn magenta (purple)
• the cause can be seen by pressing the
info
key
• the calibration data can be seen in the Diagnostics menu