![Knick Stratos Pro A201COND Скачать руководство пользователя страница 106](http://html.mh-extra.com/html/knick/stratos-pro-a201cond/stratos-pro-a201cond_user-manual_1980116106.webp)
106
Menu item
Remark
Displaying currently measured values
(sensor monitor) with HOLD mode activated:
Select MONITOR using
, press
enter
to confirm.
Select variable in the bottom text line using
.
The selected parameter is shown in the upper
display line.
As the device is in HOLD mode, you can perform
validations using simulators without influencing the
signal outputs.
Return to Service menu:
Hold
meas
depressed for longer than 2 sec.
Press
meas
once more to return to measurement.
Specifying the current at outputs 1 and 2:
Select OUT1 or OUT2 using the
keys,
press
enter
to confirm.
Enter a valid current value for the respective output
using
keys.
Press
enter
to confirm.
For checking purposes, the actual output current is
shown in the bottom right corner of the display.
End by pressing
enter
or
meas
.
OUT2:
Only if equipped with 2nd current output.