5.7.1 Sensor Data
Memosens Sensors
Memosens sensors provide relevant sensor data automatically.
Analog Sensors
The sensor type must be selected if using analog sensors:
Parameter Setting
[II] Analog …
Sensor Data
Sensor Data
Sensor Type
Standard
On
Back
Sensor Monitoring Details
Temperature Detection
Sensoface
II
Back to Meas.
01.
In
Temperature Detection
, select the used temperature probe and whether the temperature is to
be measured automatically or manually during measurement and/or calibration.
Temperature Detection (Admin.)
Measuring Temp
Back
Temperature Probe
Cal Temperature
Pt 1000
Automatic
Automatic
II
Pt 100
Pt 1000
NTC 30k
NTC 8.55k
Balco 3 kΩ
Sensoface
The Sensoface icons provide the user with diagnostic information on wear and required maintenance
of the sensor. In measuring mode, an icon (happy, neutral, or sad smiley) is shown on the display to re-
flect the continuous monitoring of the sensor parameters.
You can configure the current outputs so that a Sensoface message generates a 22-mA error signal.
Parameter Setting
Inputs and Outputs
Current Outputs
Current Output I…
Behavior during Messages
The Sensoface message can also be output via a relay contact:
Parameter Setting
Inputs and Outputs
Relay Contacts
Contact K…
Sensoface monitors the pH sensor on the basis of the following parameters:
Slope, zero point, glass impedance (if Sensocheck is enabled), response time, calibration timer, wear
Enabling/Disabling Sensoface
Sensoface is enabled and disabled in the
Sensor Data
submenu:
Parameter Setting
[I] [II] [Sensor]
Sensor Data
Note:
After a calibration, a smiley is always displayed for confirmation, even if Sensoface is disabled.