95
Menu item
Action
Select
Output current
curve
Select using
keys, confirm
by pressing enter
LIN
Linear characteristic
biLIN
Bilinear curve
LOG
Logarithmic curve
Current start
and current end
Enter value using
keys.
Press enter to confirm.
Entered value applies to
selected process variable/range.
If the adjusted range is
exceeded, the device automati-
cally switches to the next higher
range (Autorange)
Bilinear curve:
Vertex X/Y
Enter value using
keys.
Press enter to confirm.
Entered value applies to
selected vertex of bilinear curve
“Corner X“ (process variable)
and “Corner Y” (output current)
– see figure below.
Vertex of bilinear curve
Output current
Process variable
Example:
Current range 4 ... 20 mA,
Current start: 0 µS/cm
Current end: 200 µS/cm
Vertex:
“CORNER X“: 10 µS/cm (process variable)
“CORNER Y“: 12 mA (output current)
Result: The output current change in the range 0 ... 10 µS/cm
is much greater than in the range 10 ... 200 µS/cm.
Configuring the Current Output