Preparing the Shield Connection
Pre-assembled special cable for SE 655 / SE 656 sensors
Cable Preparation SE 655 / SE 656
The pre-assembled special cable:
BU
RD
RD
BU
GN
WH
YE
GN/YE
Coax RD
Coax WH
• Insert the special cable through the cable
entry into the terminal compartment.
• Remove the already separated part of the
cable insulation (1).
• Turn the shielding mesh (2) over the cable
insulation (3).
• Then shift the crimp ring (4) over the
shielding mesh and tighten it using a pince
(5)