152
Service
In the Service mode you can access the following menus:
MONITOR
Displaying currently measured values
NEW EXCHANGER Resetting the consumption calculation after replacement of
ion exchanger
POWER OUT
Power output (adjustable: 3.1/12/15/24 V)
OUT1
Testing current output 1
OUT2
Testing current output 2
RELAIS
Testing the function of the 4 relays
CONTROL
Testing the controller function
CODES
Assigning and editing passcodes
DEVICE TYPE
Selecting the measuring function
DEFAULT
Resetting the device to factory settings
OPTION
Enabling options via TAN
Note:
HOLD is active during Service mode!
Action
Key/Display
Remark
Activate Service
Press menu key to call the selection
menu.
Select SERVICE using
keys,
press enter to confirm.
Passcode
Enter passcode “5555“ for service
mode using the
keys.
Press enter to confirm.
Display
In service mode the following icons
are displayed:
• HOLD triangle
• Service (wrench)
Exit
Exit by pressing meas.