SensoGate WA131MH
33
6.3.4 Immersion Tube: Removal
1
2
3
4
01.
Remove the drive unit
(1)
.
Drive Unit: Disassembly, p. 31
02.
Move the drive unit
(1)
to the PROCESS position (sensor must be mounted).
Moving into the PROCESS Position, p. 23
03.
Loosen the screws
(2)
around 4 rotations using a screwdriver of type TX25
(3)
(do not completely
unscrew).
04.
Rotate the immersion tube
(4)
around 60° counterclockwise until the bayonet coupling of the im-
mersion tube
(4)
is open.
4
5
6
05.
Pull the immersion tube
(4)
off the sensor
(6)
.
✓
The O-ring
(5)
is now visible, or it may be located in the removed immersion tube
(4)
.
06.
Check the O-ring
(5)
for damage; replace the O-ring
(5)
if necessary.