10

9

8

11.

Position the ext

ension 

(5)

 and rotate clockwise until the bayonet coupling engages.

Cont

our 

(13)

 flush with the mark

ing 

(12)

.

12.

On first-time installation: Mount the split red servic

e cap 

(14)

 on the sensor cable 

(8)

.

13.

On first-time installation: Push the servic

e cap 

(14)

 toward the extension 

(5)

 until it positively

engages.

14.

On first-time installa

tion: Hold the sensor cable 

(8)

 in a loop and fast

en it with the clamp 

(11)

. Dur-

ing this process, the sensor cable loop must be long enough so that the sensor cable does not im-

pede the stroke movement of the SensoGate WA131M.

15.

On first-time installation: Optionally connect the equipotential bonding line

 

(10)

 t

o the clamp 

(9)

.

16.

Optional: Install ZU0759 protective cap. 

Accessories, p. 48

5.3.5  Solid-Electrolyte Sensor, Long Immersion Depth: Removal

Note: 

On versions with rinse connection, rinse the sensor prior to removal in order to prevent entrain-

ment of chemically aggressive process medium in the area of the sensor holders.

1

2

3

1

2

01.

Move the SensoGate WA131M into the SERVICE position. 

Moving into the SERVICE Position, p. 26

02.

Check outlet and leakage bores for escaping process medium. If process medium escapes, depres-
surize the process and perform troubleshooting. 

Troubleshooting, p. 42

03.

Optional: Remove ZU0759 protective cap.

Note: 

The extension can only be unlocked in the SERVICE position. The red service cap 

(2)

 must be

visible in order to unlock. 

Limit Positions, p. 20

04.

Rotate the ext

ension 

(1)

 counterclockwise until its bayonet coupling 

(1)

 unlocks.

05.

Move the extension 

(1)

 in the direction of the arrow until the cable bushing 

(3)

 is accessible.

Содержание SensoGate WA131M

Страница 1: ...User Manual SensoGate WA131M Manual Retractable Fitting www knick de Read before installation Keep for future use ...

Страница 2: ... be sufficiently reduced This document uses the following warnings to indicate hazardous situations Symbol Category Meaning Remark WARNING Designates a situation that can lead to death or serious irreversible injury The warnings contain information on how to avoid the hazard CAUTION Designates a situation that can lead to slight or moderate reversible injury Without NOTICE Designates a situation t...

Страница 3: ...0 2 Product 11 2 1 Package Contents 11 2 2 Product Identification 11 2 2 1 Example of a Version 11 2 2 2 Product Code 12 2 3 Nameplates 14 2 4 Symbols and Markings 16 2 5 Design and Function 16 2 5 1 Retractable Fitting 17 2 5 2 Drives and Sensor Holders 18 2 5 3 Process Connections 18 2 5 4 Immersion Tubes 19 2 6 Permissible Changes 19 2 7 Limit Positions 20 2 7 1 SERVICE and PROCESS Position 20 ...

Страница 4: ...mersion Lock Without a Mounted Liquid Electrolyte Sensor Functional Test 34 6 2 Preventive Maintenance 35 6 2 1 Approved Lubricants 35 6 2 2 Characteristics of Process Wetted Materials 35 6 3 Corrective Maintenance 36 6 3 1 Corrective Maintenance Safety Instructions 36 6 3 2 Drive Unit Disassembly 36 6 3 3 Drive Unit Assembly 37 6 3 4 Immersion Tube Removal 38 6 3 5 Immersion Tube Installation 39 ...

Страница 5: ... the environment The SensoGate WA131M X version is certified for operation in explosive atmospheres Operation in Explosive Atmospheres p 9 1 2 Personnel Requirements The personnel must be authorized by the operating company and instructed in handling this product The operating company must ensure that personnel are sufficiently qualified in accordance with the local and national codes and regulati...

Страница 6: ...ided on a safety label The safety label is attached to the strain relief bracket of the SensoGate WA131M 1 2 Immersion Lock Without a Mounted Liquid Electrolyte Sensor The safeguard is only available with special version V Product Code p 12 The immersion lock can be seen at the yellow indicator ring 2 above the bellows 1 If the yellow indicator ring 2 is missing the safeguard function is not avail...

Страница 7: ...ng aggressive chemical process media adjust the inspection and maintenance intervals accordingly Inspection p 33 Adhering and sticky process media can impact the functionality of SensoGate WA131M e g by causing components to stick together Adjust the inspection and maintenance intervals accord ingly Inspection p 33 1 4 2 Accidental Loosening of the Process Connection The coupling nut of the screw ...

Страница 8: ...g lug on the retainer clamp engages in the groove of the coupling nut form fit ZU0877 Locking Clamp for Process Connection G1 G1 R1 R1 1 NPT The locking clamp prevents the process screw joint of the installed SensoGate WA131M with threaded connection from accidentally loosening The locking clamp is available for process connections with the following threads G1 G1 R1 R1 1 NPT The locking clamp can...

Страница 9: ...t this deviation Nameplates p 14 Electrostatic charging The drive unit of specific versions of the SensoGate WA131M X contains housing components made of non conductive plastic Due to their surface the housing components may build up an electro static charge To prevent this charge from becoming an effective ignition source in Zone 0 ensure that the following conditions are met Highly efficient cha...

Страница 10: ...ve the necessary properties to ensure trouble free operation of the SensoGate WA131M Special applications or upgrades to special lubricants are available on request Preventive Maintenance p 35 Tools and Mounting Aids Special and accessory tools help maintenance personnel to replace components and wear parts safely and professionally Tools p 51 Spare Parts Genuine Knick spare parts are available fo...

Страница 11: ...14 2 2 1 Example of a Version Basic device with manual drive stainless steel WA131M X 0 K B H 0 A A 2 2 0 0 0 Explosion protection ATEX Zone 0 X Sensor Sensor Ø 12 mm with PG 13 5 0 Gasket material FFKM K Wetted materials Hastelloy Hastelloy Hastelloy B Process connections Ingold socket 25 mm H 0 Immersion depth Short A Electrical limit signal Without A Rinse media connection Inlet G1 8 female and...

Страница 12: ...ose 1 4571 PN10 16 DN 32 B 0 Flange loose 1 4571 PN10 16 DN 40 B A Flange loose 1 4571 PN10 16 DN 50 B 1 Flange loose 1 4571 PN10 16 DN 65 B 2 Flange loose 1 4571 PN10 16 DN 80 B 3 Flange loose 1 4571 PN10 16 DN 100 B 4 Flange loose 1 4571 PN40 DN 25 E X Flange loose 1 4571 PN40 DN 32 E 0 Flange loose 1 4571 PN40 DN 40 E A Flange loose 1 4571 PN40 DN 50 E 1 Flange loose 1 4571 PN40 DN 65 E 2 Flang...

Страница 13: ...l Without A With B Rinse media connection Without inlet outlet G1 8 female with outlet hose complete 3 m 0 Inlet G1 8 female outlet G1 8 female with outlet hose complete 3 m 1 Inlet G1 8 female and inlet hose complete 5 m outlet G1 8 female with outlet hose complete 3 m 2 Housing material Stainless steel PP 6 bar operating pressure 1 Stainless steel PEEK 10 bar operating pressure 2 Special version...

Страница 14: ...rsion of the SensoGate WA131M Nameplate Version With Ex Approval Note The figure shows a nameplate for the SensoGate WA131M X version by way of example Max pressure Temperature range See Drive Unit for Ex marking 0044 SensoGate Process Unit 14163 Berlin Made in Germany Type WA131M X KBH0 2 000 No 2203709 1906 WA131M X0K AA 2 000 2203711 1910 1 2 1 Drive unit nameplate 2 Process unit nameplate ...

Страница 15: ...te The figure shows a nameplate for the SensoGate WA131M N version by way of example 14163 Berlin Made in Germany SensoGate Type No Process Unit WA131M N0A AA 1 000 2203710 1907 WA131M N AAN1 1 000 2203708 1906 1 2 1 Drive unit nameplate 2 Process unit nameplate ...

Страница 16: ...ve the SensoGate WA131M to the PROCESS position Moving into the PROCESS Position p 25 Symbol indicating the direction of rotation to move the SensoGate WA131M to the SERVICE position Moving into the SERVICE Position p 26 Outlet symbol marking the outlet port of the SensoGate WA131M Inlet symbol marking the inlet port of the SensoGate WA131M 1 2 5 Design and Function SensoGate WA131M consists of tw...

Страница 17: ...it 8 Inlet port1 2 Process unit 9 Immersion tube 3 Protection sleeve 10 Process connection e g Ingold socket 4 Strain relief bracket with grounding connection 11 Calibration chamber base 5 Rotating collar 12 Calibration chamber top 6 Coupling nut 13 Leakage bore 7 Outlet port 14 Safety lock button 1 Availability is dependent on the ordered version Product Code p 12 ...

Страница 18: ...ve short ID 1 solid electrolyte sensor 225 mm 3 Drive short ID 1 liquid electrolyte sensor 250 mm 2 Drive long ID 1 solid electrolyte sensor 225 mm 2 5 3 Process Connections Note The figure shows a selection from the product line Product Code p 12 1 2 3 4 1 Flange 3 Dairy pipe screw joint 2 Ingold socket 4 1 NPT male 1 ID immersion depth ...

Страница 19: ... the sensor holder to fit another sensor type Drives and Sensor Holders p 18 Retrofit of safeguards e g Immersion lock without a mounted liquid electrolyte sensor Safeguards p 6 Any changes may result in deviations between the information on the nameplate and the actual ver sion of the SensoGate WA131M The operating company is responsible for evaluating the permissibil ity of the changes and for d...

Страница 20: ...sed Solid electrolyte sensor short immersion depth Solid electrolyte sensor long immersion depth Liquid electrolyte sensor short immersion depth 1 2 3 1 2 4 1 2 5 SERVICE PROCESS SERVICE PROCESS SERVICE PROCESS In the SERVICE position the sensor head 3 is visible at the top end of the protection sleeve In the SERVICE position the service cap 4 is visible at the top end of the extension In the SERV...

Страница 21: ...tch 1 has the following characteristics Need not be marked according to EN 60079 For connection to intrinsically safe circuits only Connection and ambient conditions Ui 30 V Ii 100 mA Pi 750 mW Ci negligibly low Li negligibly low Temperature class T6 T6 T5 T5 Equipment protection level Ga Gb Ga Gb Ambient temperature range 10 C 45 C 14 F 113 F 10 C 60 C 14 F 140 F 10 C 57 C 14 F 134 6 F 10 C 70 C ...

Страница 22: ...ope of delivery of the SensoGate WA131M for completeness Package Contents p 11 02 Check the SensoGate WA131M for damage 03 Ensure the required sensor installation clearances Dimension Drawings p 52 04 Fasten the SensoGate WA131M to the process port using the process connection 05 Optional If using the product in explosive atmospheres connect the grounding connection of SensoGate WA131M to the plan...

Страница 23: ... the hose clamp 3 3 4 Inlet Hose Installation NOTICE Drinking water may be contaminated by rinse and process media when connecting to drink ing water pipes Observe the information contained in EN 1717 Install a suitable check valve e g check valve RV01 at the water or rinse connection Accessories p 48 1 2 5 4 3 1 5 4 3 Note When using versions of SensoGate WA131M with inlet port the sealing insert...

Страница 24: ...emoving Sensors p 27 05 Ensure that the process connection is securely fastened 06 Optional Ensure that installed safety accessories e g ZU0818 retainer clamp are securely fas tened Safety Accessories p 8 07 Optional Ensure that the SensoGate WA131M X is correctly connected to the plant s equipotential bonding system Operation in Explosive Atmospheres p 9 08 Move the SensoGate WA131M into the PROC...

Страница 25: ...ety lock button has popped out is the function of the safeguard Immersion lock without a mounted sensor ensured Safeguards p 6 SERVICE PROCESS 1 2 A B 01 Mount the sensor Installing and Removing Sensors p 27 Note When the rotary movement starts the safety lock button is automatically depressed 02 Depress the safety lock button 2 see detail A and rotate the rotating collar 1 clockwise The safety lo...

Страница 26: ...l B Only if the safety lock button has popped out is the function of the safeguard Immersion lock without a mounted sensor ensured Safeguards p 6 PROCESS SERVICE 1 2 A B Note When the rotary movement starts the safety lock button is automatically depressed 01 Depress the safety lock button 2 see detail A and rotate the rotating collar 1 counterclockwise The safety lock button 2 pops out when the S...

Страница 27: ...splash out during sensor re placement 5 3 2 Solid Electrolyte Sensor Short Immersion Depth Installation 1 2 3 10 9 3 7 8 6 5 4 01 Move the SensoGate WA131M into the SERVICE position Moving into the SERVICE Position p 26 02 Check outlet and leakage bores for escaping process medium If process medium escapes depres surize the process and perform troubleshooting Troubleshooting p 42 03 Check the comp...

Страница 28: ...r to prevent entrain ment of chemically aggressive process medium in the area of the sensor holders 1 2 3 5 4 01 Move the SensoGate WA131M into the SERVICE position Moving into the SERVICE Position p 26 02 Check outlet and leakage bores for escaping process medium If process medium escapes depres surize the process and perform troubleshooting Troubleshooting p 42 03 Optional Remove ZU0759 protecti...

Страница 29: ...f necessary 04 Rotate the extension 5 counterclockwise until its bayonet coupling 5 unlocks 05 Move the extension 5 in the direction of the arrow and remove it 1 4 4 5 6 7 8 06 Push in the sensor 4 Note When tightening the sensor the spring force of the Immersion lock without a mounted solid electrolyte sensor safeguard must be overcome 07 Tighten the sensor 4 using the spanning wrench 1 to max 3 ...

Страница 30: ...al Install ZU0759 protective cap Accessories p 48 5 3 5 Solid Electrolyte Sensor Long Immersion Depth Removal Note On versions with rinse connection rinse the sensor prior to removal in order to prevent entrain ment of chemically aggressive process medium in the area of the sensor holders 1 2 3 1 2 01 Move the SensoGate WA131M into the SERVICE position Moving into the SERVICE Position p 26 02 Chec...

Страница 31: ... 4 01 Move the SensoGate WA131M into the SERVICE position Moving into the SERVICE Position p 26 02 Check outlet and leakage bores for escaping process medium If process medium escapes depres surize the process and perform troubleshooting Troubleshooting p 42 03 Loosen the small coupling nut 1 by a few rotations do not loosen completely 04 Fully loosen the large coupling nut 2 and pull off the enti...

Страница 32: ...l in order to prevent entrain ment of chemically aggressive process medium in the area of the sensor holders 1 2 3 01 Move the SensoGate WA131M into the SERVICE position Moving into the SERVICE Position p 26 02 Check outlet and leakage bores for escaping process medium If process medium escapes depres surize the process and perform troubleshooting Troubleshooting p 42 03 Disconnect the sensor cabl...

Страница 33: ... O rings and re grease Corrective Maintenance p 36 6 1 2 Knick Premium Service Knick offers individually compiled services tailored to the customer s requirements for inspections and functional tests on the product Further information can be found at www knick de 6 1 3 Immersion Lock Without a Mounted Solid Electrolyte Sensor Functional Test 01 Move the SensoGate WA131M into the SERVICE position M...

Страница 34: ...ressure resistance is still available in the event of a malfunction 03 Loosen the large coupling nut 3 around 1 5 rotations do not loosen completely 04 Check the function of the Immersion lock without a mounted liquid electrolyte sensor It must be impossible to depress the safety lock button It must be impossible to rotate the rotating collar 05 Fasten the large coupling nut 3 finger tight 06 Fast...

Страница 35: ...stated values A pretest is recom mended for cases where there has been no prior experience using the material in the process Mix tures of substances constitute a prime example Mechanical strength Temperature resistance Resistance to acids Resistance to alkalis Resistance to salt solutions Resistance to cleaning agents or solvents Stainless steel material no 1 4571 1 1 3 3 2 3 2 Hastelloy C 22 mate...

Страница 36: ...ect the SensoGate WA131M from the process Retractable Fitting Removal p 44 02 As necessary disconnect the outlet hose inlet hose1 and limit switch1 03 Move the SensoGate WA131M into the SERVICE position Moving into the SERVICE Position p 26 04 As necessary remove the sensor 3 Installing and Removing Sensors p 27 Note Do not tilt the coupling nut Use a suitable spanning wrench e g the one contained...

Страница 37: ...s unit 5 While doing so posi tion the coding pin 4 in the hole 2 Note Do not tilt the coupling nut Use a suitable spanning wrench e g the one contained in ZU0680 service set or ZU0740 service set Tools p 51 03 Position the coupling nut 8 and tighten clockwise finger tight or to 10 Nm using the spanning wrench 6 04 As required install the outlet hose at the outlet 7 Outlet Hose Installation p 23 05...

Страница 38: ...en the screws 2 around 4 rotations using a screwdriver of type TX25 3 do not completely unscrew 04 Rotate the immersion tube 4 around 60 counterclockwise until the bayonet coupling of the im mersion tube 4 is open 4 5 6 05 Pull the immersion tube 4 off the sensor 6 The O ring 5 is now visible or it may be located in the removed immersion tube 4 06 Check the O ring 5 for damage replace the O ring 5...

Страница 39: ...g a screwdriver of type TX25 6 do not completely unscrew Note There may be an O ring in the immersion tube left over from the removal process Remove this O ring prior to removal of the immersion tube 06 Carefully push the immersion tube 4 onto the sensor 3 and insert it into the bayonet cou pling 5 1 4 6 07 Firmly push the immersion tube 4 into the bayonet coupling 5 at the same time rotating arou...

Страница 40: ...cess unit from the drive unit Drive Unit Disassembly p 36 02 Remove the screws 1 with a screwdriver of type TX25 Keep the screws 1 in a safe place for as sembly later on 03 Position the pliers 3 and use the face pin spanner wrench 2 to loosen the coupling of the split calibration chamber 4 5 04 Unscrew the top 4 from the bottom 5 of the calibration chamber and separate the two parts ...

Страница 41: ...ts p 45 02 Connect the top 1 and the bottom 2 of the calibration chamber and screw together finger tight 03 Position the pliers 4 and use the face pin spanner wrench 3 to screw the calibration chamber together 5 5 Note Securing the calibration chamber with the two screws is not possible until the top and bottom parts have been firmly screwed together to the hard stop 04 Tighten the screws 5 with a...

Страница 42: ...e Fitting Emergency Release p 43 Sensor glass shattered Mechanical impact on the sensor glass e g by process medium Replace faulty sensor Installing and Removing Sensors p 27 Remove any glass splinters from the SensoGate WA131M Check immersion tube seal and replace if necessary Gasket Sets p 45 No or wrong measured value displayed Sensor may be faulty Replace the sensor Installing and Removing Sen...

Страница 43: ...h the set screw 4 Drive Unit Disassembly p 36 02 Using the Allen wrench A F 2 5 mm 1 screw in the set screw 4 up to the stop 03 Move the SensoGate WA131M into the SERVICE position Moving into the SERVICE Position p 26 04 Perform troubleshooting as required send the SensoGate WA131M to Knick for repair Malfunction States p 42 Note The function of the safeguard Immersion lock without a mounted senso...

Страница 44: ...8 retainer clamp 06 Loosen the process connection 07 Remove the SensoGate WA131M from the customer s process port 08 Close off the process port appropriately 8 2 Returns If required send the SensoGate WA131M in a clean condition and securely packed to Knick Elektronische Messgeräte GmbH Co KG If the SensoGate WA131M has been in contact with hazardous substances it must be decontaminated disinfecte...

Страница 45: ...KM ZU0689 1 Set A 2 Process wetted gasket material FKM wetted by rinse medium FKM ZU0829 Set B 1 Process wetted gasket material EPDM ZU0690 1 Set B 2 Process wetted gasket material EPDM wetted by rinse medium EPDM ZU0830 Set E 1 Process wetted gasket material EPDM FDA ZU0692 1 Set E 2 Process wetted gasket material EPDM FDA wetted by rinse medium EPDM FDA ZU0831 Set K 1 Process wetted gasket mater...

Страница 46: ...ed O rings Rinse wetted O rings Scraper ring 215 000 420 23x2 11 9x2 6 20x2 5 20x2 5 215 000 420 23x2 13x1 5 8x1 5 40x2 5 Gasket sets for socket process connection Process wetted O rings Rinse wetted O rings Scraper ring 215 000 420 23x2 11 9x2 6 20x2 5 20x2 5 21x2 20x2 13x1 5 8x1 5 40x2 5 ...

Страница 47: ...04 mm Materials ZU0826 PP ZU0725 PEEK HD ZU0727 PVDF HD ZU0739 Bellows The bellows only used on versions with liquid electrolyte sensors protect the fit ting beneath the pressure chamber against external contamination and wear ZU0889 Outlet Hose The outlet hose is used to drain calibration cleaning or rinse media from the cali bration chamber Outlet Hose Installation p 23 ZU0760 Scraper Ring Reinf...

Страница 48: ...lectrolyte sensors ZU0759 1 suitable for versions with liquid electrolyte sensors ZU0717 Straight Weld In Socket for Tank Walls Process connection Ingold socket Ø 25 mm G1 ZU0717 DN Straight Weld In Socket for Pipelines Process connection Ingold socket Ø 25 mm G1 adapted to DN50 ZU0717 DN50 adapted to DN65 ZU0717 DN65 adapted to DN80 ZU0717 DN80 adapted to DN100 ZU0717 DN100 ZU0718 15 Incline Weld...

Страница 49: ...ld In Socket HSD for Piping Process connection Ingold socket Ø 25 mm G1 adapted to DN50 ZU0922 DN50 adapted to DN65 ZU0922 DN65 adapted to DN80 ZU0922 DN80 adapted to DN100 ZU0922 DN100 ZU0923 15 Incline Safety Weld In Socket HSD for Tank Walls Process connection Ingold socket Ø 25 mm G1 ZU0923 DN 15 Incline Safety Weld In Socket HSD for Piping Process connection Ingold socket Ø 25 mm G1 adapted t...

Страница 50: ...1 NPT The locking clamp prevents the process screw joint of the installed SensoGate WA131M with threaded connection from accidentally loosening The locking clamp is available for process connections with the following threads G1 G1 R1 R1 1 NPT The locking clamp can be used with threaded couplings with a minimum length of 10 mm and an outer diameter of 39 mm to 57 mm Flange Protector The flange pro...

Страница 51: ...ive maintenance and corrective maintenance as well as to modify the product SensoGate WA131M can be fully dismantled using this tool set ZU0754 SensoGate Service Set Calibration Chamber This tool set is suitable for maintenance work on the calibration chamber and its seals It allows easy separation of the split calibration chamber ZU0746 Accessory Tool for Scraper Ring The ZU0746 accessory tool al...

Страница 52: ...SensoGate WA131M 52 10 Dimension Drawings Retractable fitting for solid electrolyte sensor short immersion depth Note All dimensions are given in millimeters inches ...

Страница 53: ...SensoGate WA131M 53 Retractable fitting for solid electrolyte sensor long immersion depth Note All dimensions are given in millimeters inches ...

Страница 54: ...SensoGate WA131M 54 Retractable fitting for liquid electrolyte sensor short and long immersion depth Note All dimensions are given in millimeters inches ...

Страница 55: ...DN 100 Flange loose ANSI 316 150 lbs 1 1 2 4 Flange loose ANSI 316 300 lbs 1 1 2 3 Short and long immersion depth ID 65 2 56 short ID 88 3 46 short ID 120 4 72 long ID 143 5 63 long ID Dairy pipe DN50 DN100 Short and long immersion depth ID 71 2 80 short ID 94 3 70 short ID 126 4 96 long ID 149 5 87 long ID 25 mm Ingold socket Short and long immersion depth ID ...

Страница 56: ... short ID 61 2 40 short ID G1 1 2 20 0 79 116 4 57 long ID 139 5 47 long ID SW55 A F 2 17 G1 outer Short and long immersion depth ID 84 3 31 short ID 61 2 40 short ID G1 1 4 20 0 79 116 4 57 long ID 139 5 47 long ID SW49 A F 1 93 G1 outer Short and long immersion depth ID ...

Страница 57: ... 3 94 short ID 132 5 2 long ID 155 6 10 long ID Clamp 1 5 and clamp 2 Short and long immersion depth ID 20 0 79 61 2 40 short ID 84 3 31 short ID NPT1 A F 1 11 16 116 4 57 long ID 139 5 47 long ID 1 NPT outer Short and long immersion depth ID ...

Страница 58: ...ic in SERVICE Position 16 bar at 0 40 C 230 psi 32 104 F 10 bar at 5 20 C 150 psi 41 68 F PP Permissible rinsing pressure and temperature 6 bar at 5 90 C 90 psi 41 194 F Ambient temperature 10 70 C 14 158 F Degree of protection IP66 Housing material Stainless steel PEEK PP EPDM Duran Sensors Product Code p 12 Process Connections Product Code p 12 Connections Inlet Female thread G 1 8 Outlet Female...

Страница 59: ...SensoGate WA131M 59 Appendix Return Form ...

Страница 60: ...luded accessories The product has not been exposed to hazardous substances or mixtures The product has been exposed to hazardous substances or mixtures State the classification of the hazardous substance as applicable together with the hazard statements or R phrases or at minimum provide the relevant hazard pictograms The product has been exposed to infectious substances The product was subjected ...

Страница 61: ...g the actual condition of an item under review including determining the causes of wear and deriving the necessary steps for future use Maintenance Combination of all technical administrative and managerial actions during the life cycle of an item intended to retain it in or restore it to a state in which it can perform the required function Source EN 13306 Maintenance Maintenance terminology Prev...

Страница 62: ...ironmental factors 7 Equipotential bonding Avoidance of possible ignition hazards 9 Connection 28 Equipotential bonding line 30 Error elimination 42 Explosive atmospheres 9 F First inspection 33 Functional description retractable fitting 16 Functional test Immersion lock w o a liquid electrolyte sensor 34 Immersion lock w o a solid electrolyte sensor 33 G Gasket material 12 Gasket sets 45 Genuine ...

Страница 63: ...l 5 R Remedies malfunctions 42 Removal from operation 44 Removal retractable fitting 44 Residual risks 7 Retractable fitting Changes 19 Function 16 Installation angle 22 Leaking 42 Main assemblies 16 Return form 44 Returns 44 Rinsing pressure permissible 58 Risk assessment ii 8 S Safeguards Overview 6 Retrofit 19 Safety accessories Locking clamp 8 Retainer clamp K8 8 Retainer clamp Ingold socket 2...

Страница 64: ...e latest documents are available for download on our website below the corresponding product description TA 215 302 KNEN07 097300 Knick Elektronische Messgeräte GmbH Co KG Headquarters Beuckestraße 22 14163 Berlin Germany Phone 49 30 80191 0 Fax 49 30 80191 200 info knick de www knick de Local Contacts www knick international com ...