Position the ext
ension
(5)
and rotate clockwise until the bayonet coupling engages.
✓
Cont
our
(13)
flush with the mark
ing
(12)
.
12.
On first-time installation: Mount the split red servic
e cap
(14)
on the sensor cable
(8)
.
13.
On first-time installation: Push the servic
e cap
(14)
toward the extension
(5)
until it positively
engages.
14.
On first-time installa
tion: Hold the sensor cable
(8)
in a loop and fast
en it with the clamp
(11)
. Dur-
ing this process, the sensor cable loop must be long enough so that the sensor cable does not im-
pede the stroke movement of the SensoGate WA131M.
15.
On first-time installation: Optionally connect the equipotential bonding line
(10)
t
o the clamp
(9)
.
16.
Optional: Install ZU0759 protective cap.
5.3.5 Solid-Electrolyte Sensor, Long Immersion Depth: Removal
Note:
On versions with rinse connection, rinse the sensor prior to removal in order to prevent entrain-
ment of chemically aggressive process medium in the area of the sensor holders.
1
2
3
1
2
01.
Move the SensoGate WA131M into the SERVICE position.
Moving into the SERVICE Position, p. 26
02.
Check outlet and leakage bores for escaping process medium. If process medium escapes, depres-
surize the process and perform troubleshooting.
03.
Optional: Remove ZU0759 protective cap.
Note:
The extension can only be unlocked in the SERVICE position. The red service cap
(2)
must be
visible in order to unlock.
04.
Rotate the ext
ension
(1)
counterclockwise until its bayonet coupling
(1)
unlocks.
05.
Move the extension
(1)
in the direction of the arrow until the cable bushing
(3)
is accessible.