36
Menu Display
Maintenance
Opening the Maintenance Menu
From the measuring mode:
Press
menu
key to select menu.
Select Maintenance (maint) using
arrow keys, press
enter
to confirm.
Passcode 2958 (The passcode can be
edited by the administrator.)
Then select "Sensor SE 740".
Sensor Monitor
You can use the sensor monitor to
validate the sensor by immersing
it in a solution with known oxygen
concentration, for example, and then
checking the values measured.
Menu selection
Return to meas
25.6°C
Select:
[enter]
Lingua
83.3 %Air
Maintenance
Sensor monitor, autoclaving counter
Note:
Function check (HOLD) mode active (Setting: BASE module)
Sensor SE 740
25.6°C
83.3 %Air
Sensor monitor
Autoclaving counter
Maintenance
25.6°C
83.3 %Air
Sensor SE 740
Module LDO 3400-170
Module BASE 3400-029
Sensor SE 740
25.6°C
83.3 %Air
Sensor monitor
Autoclaving counter
Autoclaving Counter
After having completed an auto-
claving process, you must manually
increment the autoclaving count.
N
ot
e:
T
he displa
y ma
y v
ar
y depending on the devic
e v
ersion.