63
• Select Curve,
e.g. “linear”: The measured variable is
represented by a linear output cur-
rent curve. The desired range of the
measured variable is specified by the
values for “Start” and “End”.
Current Outputs
Select menu: Parameter setting/Module BASE
Note: Function check (HOLD) mode active
Menu Display
Parameter setting
BASE module
• Select process variable
Output current
[pH]
20
4
14
0
Assigning measured values: Start (4 mA) and End (20 mA)
[pH]
20
4
7
5
[mA]
Example 1: Range pH 0 - 14
Example 2: Range pH 5 - 7
Advantage: Higher resolution in range of interest
[mA]
To configure current output
• Call up parameter setting
• Enter passcode
• Select “Module BASE”
• Select “Output current I1”
Configuring a Current Output
• Open parameter setting
• Enter passcode
• Select “Module BASE”
• Select “Output current ...”
Module BASE (Administrator)
19.2°C
Return
Contact K2 (NAMUR HOLD)
Contact K4 (NAMUR Failure)
Contact K3 (NAMUR maintenance)
7.00 pH
Output current I2
Contact K1 (Limit)
Inputs OK1, OK2
Block
Output current I1 (Administrator)
OK
19.0 °C
7.00 pH
Variable
Output
Abort
Curve
Linear
°C
Trilinear
Function
Table
Output current I1 (Administrator)
OK
19.0 °C
7.00 pH
Curve
Output
Behavior during messages
Start
Abort
Variable
°C
rH
S/cm
°C
g/kg
Off
pH
End
Start
End
Output filter
5 7
N
ot
e:
T
he displa
y ma
y v
ar
y depending on the devic
e v
ersion.
Содержание Protos II 4400
Страница 18: ...18 ...
Страница 23: ...23 Wiring Examples Connection of ISM sensor Transparent Black PH 3400 X 035 ...
Страница 73: ...73 ...
Страница 88: ...88 Appendix Buffer table Mettler Toledo ...
Страница 90: ...90 Appendix Buffer table DIN 19267 ...
Страница 92: ...92 Appendix Buffer table Techn buffers to NIST ...
Страница 93: ...93 Appendix Buffer table Hamilton ...
Страница 94: ...94 Appendix Buffer table Kraft ...
Страница 95: ...95 Appendix Buffer table Hamilton A ...
Страница 96: ...96 Appendix Buffer table Hamilton B ...
Страница 97: ...97 Appendix Buffer table HACH ...
Страница 100: ...100 ...
Страница 111: ...111 Index Protos PH 3400 X 035 Module ...