![Knick Protos CONDI 3400(X)-051 Скачать руководство пользователя страница 59](http://html.mh-extra.com/html/knick/protos-condi-3400-x-051/protos-condi-3400-x-051_user-manual_1980169059.webp)
59
Messages
Setting the Message Parameters
Messages
Note: HOLD mode active
Menu Display
Messages
All parameters determined by the
measuring module can generate
messages.
• Device limits max:
Messages are generated when the
process variable (e.g. conductivity) is
outside the measurement range. The
“Failure” icon is displayed, the NAMUR
failure contact is activated (BASE
module, factory setting: contact K4,
N/C contact).
The current outputs can signal a
22 mA message (user defined).
• Variable limits:
For the “failure” and “warning”
messages you can define upper and
lower limits for message generation.
• Message icons:
Messages (Administrator)
20.1 °C
Return
0.220 mS/cm
Messages Conductivity (Administrator)
20.1 °C
Abort
Monitoring
0.220 mS/cm
Variable limits
OK
Device limits max.
Off
Messages Temperature
Messages Resistivity
Message Conductivity
Messages Salinity
Messages Concentration
Messages (Administrator)
20.1 °C
0.220 mS/cm
Warning Limit Hi
Monitoring
Failure Limit Lo
Failure Limit Hi
Warning Limit Lo
+1999 mS/cm
Variable limits
+1999 mS/cm
-0.000 μS/cm
-0.000 μS/cm
Abort
OK
Message list
22.3 °C
Fail Meas. processing
0.220 mS/cm
Fail Temperature range
Abort
Diagnostics menu
When the “Maintenance” or “Failure”
icons are flashing in the display, you
should open the Diagnostics menu.
The messages are displayed in the
“Message list”.
Failure (Failure limit HiHi/LoLo)
Maintenance (Warning limit Hi/Lo)
Содержание Protos CONDI 3400(X)-051
Страница 24: ...24 ...
Страница 38: ...38 ...
Страница 51: ...51 Nitric acid HNO3 Hydrochloric acid HCl c by wt c by wt ...
Страница 52: ...52 Sodium hydroxide solution NaOH Table salt solution NaCl c by wt c by wt ...