24
Commissioning and Configuration
Corresponding to example given on Page 18, "Offline Configuration"
For parameter changes, you must set MODE_BLK/TARGET to OOS (Out of
Service) since otherwise the error message NIF_ERR_WRONG_MODE_FOR_
REQUEST would appear for [Write Changes].
Analog_Input_1
"Process" card:
CHANNEL
Module 1 - Channel 1
"Scaling" card:
XD_SCALE/UNITS_INDEX
pH
OUT_SCALE/UNITS_INDEX
pH
L_TYPE
Direct
Button
[Write Changes]
Button
[Auto]
Analog_Input_2
"Process" card:
CHANNEL
Module 1 - Channel 2
"Scaling" card:
XD_SCALE/UNITS_INDEX
°C
OUT_SCALE/UNITS_INDEX
°C
L_TYPE
Direct
Button
[Write Changes]
Button
[Auto]
Analog_Input_3
"Process" card:
CHANNEL
Module 1 - Channel 3
"Scaling" card:
XD_SCALE/UNITS_INDEX
mV
OUT_SCALE/UNITS_INDEX
mV
L_TYPE
Direct
Button
[Write Changes]
Button
[Auto]
Analog_Input_4
"Process" card:
CHANNEL
Module 1 - Channel 4
"Scaling" card:
XD_SCALE/UNITS_INDEX
Mohm
OUT_SCALE/UNITS_INDEX
Mohm
L_TYPE
Direct
Button
[Write Changes]
Button
[Auto]