93
System Forecast: Tolerance Specifications
Menu: Parameter setting / Probe control / Installation / System forecast
System forecast: Tolerance specifications
Menu Display
System forecast: Tolerances
Tolerance specifications
For monitoring during operation,
permissible deviations in the sep-
arate travel times can be set in the
"Tolerances" menu. The tolerances
"a ... e" are permissible deviations
from the active reference profile.
If monitoring is set to "on", the
icon is displayed in measuring mode
if a tolerance limit is exceeded – the
corresponding message text can
be viewed under "Diagnostics /
Messages".
Return
25.6 °C
7.00 pH
12/21/16 12:37 S a06 b17 c25 d33 e02 CM
Monitoring
Edit reference profile
12/21/16 12:37 P a09 b16 c26 d34 e02 CM
Active reference profile
Tolerances
System control (Administrator)
WA150
Off
On
Tolerances (tol.)
Input range
a
Probe valve switching time (tol.+)
00 ... 99
[1/10 s]
b
Exit of limit position (tol.+)
00 ... 99
[1/10 s]
c
Achievement of limit position (tol.+) 00 ... 99
[1/10 s]
d
Air sensor deactivated (tol.+)
00 ... 99
[1/10 s]
Air sensor deactivated (tol.–)
00 ... 99
[1/10 s]
e
Air current interruptions (tol.+)
01 ... 09
[number]
Reference profile tolerance
00 ... 99 %
Содержание Protos 3400
Страница 12: ...12 ...
Страница 160: ...160 ...
Страница 177: ...177 Appendix Buffer Table Mettler Toledo ...
Страница 179: ...179 Appendix Buffer Table DIN 19267 ...
Страница 181: ...181 Appendix Buffer Table Techn Buffers to NIST ...
Страница 182: ...182 Appendix Buffer Table Hamilton ...
Страница 183: ...183 Appendix Buffer Table Kraft ...
Страница 184: ...184 Appendix Buffer Table Hamilton A ...
Страница 185: ...185 Appendix Buffer Table Hamilton B ...
Страница 186: ...186 Appendix Buffer Table HACH ...
Страница 206: ...206 Index ...
Страница 207: ...207 Index ...