62
Module BASE (Administrator)
Output current I1
Return
Block
7.00 pH
Contact K4
Output current I2
Contact K3
19.0°C
(NAMUR Failure)
Contact K2
Contact K1
(NAMUR maintenance)
( Sensoface)
( Sensoface)
Relay Contacts: Sensoface Messages
Parameter setting/Module BASE/Relay contacts/Usage/Sensoface
Set contact parameters
• (e.g. "N/O")
• Set ON / OFF delay.
Menu Display
Parameter setting (Sensoface)
Relay contacts, usage
• Open parameter setting
• Enter passcode
• Select “Module BASE”
• Select contact e.g. K1)
• Assign Sensoface message of
desired measuring module to
selected relay contact
Contact K1 (Administrator)
19.2°C
Return
Hysteresis
Variable
Limit value
7.00 pH
Usage
Effective direction
Contact type
Limit value
NAMUR maintenance
NAMUR HOLD
Sensoface
Parameter set B active
USP output
Contact K1 (Administrator)
19.2°C
Return
OFF delay
ON delay
7.00pH
Usage
Contact type
0010 s
Sensoface
N/O
0010 s
Assign Sensoface messages to relay
contacts
When more than one measuring
module is used, the Sensoface mes-
sages of the modules can be assigned
to different contacts.
Содержание Protos 3400
Страница 12: ...12 ...
Страница 160: ...160 ...
Страница 177: ...177 Appendix Buffer Table Mettler Toledo ...
Страница 179: ...179 Appendix Buffer Table DIN 19267 ...
Страница 181: ...181 Appendix Buffer Table Techn Buffers to NIST ...
Страница 182: ...182 Appendix Buffer Table Hamilton ...
Страница 183: ...183 Appendix Buffer Table Kraft ...
Страница 184: ...184 Appendix Buffer Table Hamilton A ...
Страница 185: ...185 Appendix Buffer Table Hamilton B ...
Страница 186: ...186 Appendix Buffer Table HACH ...
Страница 206: ...206 Index ...
Страница 207: ...207 Index ...