22
Inserting the Module
Note: Be sure to connect the shielding properly!
Switch off power supply
Open the device (loosen the 4 screws at the front)
Place module in slot (D-SUB connector)
Tighten fastening screws of the module
Open ESD shielding cap (covering terminals 2 and 8)
Connect sensor cable.
To avoid interferences, the cable shielding must be completely
covered by the ESD shielding cap.
Close ESD shielding cap (covering terminals 2 and 8)
Connect RS 485 interface to probe controller
(terminals 16 ... 19)
Close device, tighten screws at the front
Switch on power supply
Set parameters
Make sure that the cable
glands are tightly closed to
protect against humidity.
The terminals 2 and 8 are
covered by an ESD shield.
To connect the sensor cable,
just pull it back.
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.