53
pH
Oxy Cond
ORP
Options
User Management (Access Control)
You can create up to 4 users with different access rights for configuration or
calibration.
Enabling the User Management
1) In measuring mode, press the Menu softkey.
2) Select “Configuration” and confirm by pressing enter.
3) Select “User management” / “Management – Enable”.
4) Select
User = ADMIN
PIN code = 1989 (factory setting)
5) Press enter.
6) To create more users / assign more PIN codes: Press Continue softkey.
Continue
Return
Configuration
Management – Disable
+ User 1
+ User 2
enter
+ User 3
+ User 4
Continue
Return
Configuration
Management – Disable
- User 1
ADMIN
PIN code
cal level
conf level
1989
Access
Access
Creating a User / Changing the PIN Code
1. Select a user (e.g., “User 1”, default: ADMIN, PIN code 1989):
Continue
Return
Configuration
Management – Disable
- User 1
ADMIN
PIN code
cal level
conf level
1989
Access
Access
2. Select ADMIN to open the editor for entering the user name:
enter
Finish
abc
Configuration
End
1 2 3 4 5 6 7 8 9 0 ? `
ADMIN |
Q W E R T Z U I O P
*
A S D F G H J K L
'
> Y X C V B N M ; : _
You can allow or
block access to
configuration or
calibration for
each user.
NOTICE!
If you lose the PIN code for the ADMIN user, the system access will be blocked.
The manufacturer can generate a rescue PIN code.