17
pH
Oxy Cond
ORP
Messages
Shows all active error and status messages as well as supplementary help texts.
MemoLog (Memosens only)
Displays the individual calibration records stored in the device. You have the possibil-
ity to delete individual entries or all entries. The following parameters are displayed:
• Sensor type
• Serial no.
• TAG
• Calibration date
• Zero point
• Slope
• Cell constant (Cond sensor)
• Operating point (ISFET sensor)
Background: The device provides a calibration data logger, which must be activated
in the configuration menu. With “MemoLog” activated, up to 100 calibration records
can be directly saved to the device. After every calibration, the complete Memosens
data will be recorded. Convenient management of the calibration data is possible
using the MemoSuite or Paraly SW 112 software.
MemoLog is not suitable for SE 340 (optical oxygen sensor).
Information
Sensor Monitor
Shows the raw values available fom the connected sensor:
pH
mV, temperature, glass impedance
pH ISFET
mV, leakage current, temperature
ORP
mV, temperature
Cond
Resistance, conductance, temperature
Oxy
Sensor current, leakage current, polarization voltage,
partial pressure, air pressure, temperature
Oxy, optical
Partial pressure, temperature
Содержание Portavo 908 MULTI