pH
Oxy Cond
ORP
Specifications
Connections
2x socket, 4 mm dia., for separate temp. detector
1 x M8 socket, 4 pins, for Memosens lab cable
1x micro USB-B for data transmission to PC
1 x socket depending on device version:
Portavo 907 MULTI PH: pH socket acc. to DIN 19 262
Portavo 907 MULTI COND: Multi-contact for 2-/4-el. sensors
Portavo 907 MULTI OXY: M12, 8 pins, for Memosens sensors
or SE 340 sensor (optical oxygen)
Air pressure measurement
700 … 1100 hPa
User interface
Straightforward menu navigation with graphic icons
and detailed operating instructions in plain text
Languages
German, English, French, Spanish, Italian, Portuguese, Russian
Sensoface
Status indication (friendly, neutral, sad)
Status indicators
For battery power level, logger
Graphic display
QVGA TFT display with white backlighting
Keypad
[on/off], [meas], [enter],
,
2 context-sensitive softkeys
Data logger
10 000 memory locations
Recording
Manual, interval- or event-controlled, with management of tag num-
bers and notes
Cal data logger
MemoLog
(Memosens only)
Up to 100 Memosens calibration records can be saved
Recording
Directly retrievable via MemoSuite
or Paraly SW 112 (USB)
Viewable on the
display
Manufacturer, sensor type, serial no., zero, slope,
calibration date
Temperature input
2 x 4 mm dia. for integrated or separate temperature detector
Measuring ranges
NTC30 temperature detector
–20 … +120 °C / –4 … +248 °F
Pt1000 temperature detector –40 … +250 °C / –40 … +482 °F
Measuring cycle
Approx. 1 s
Measurement error
1,2,3)
< 0.2 K (Tamb = +23 °C / +73.4 °F); TC < 25 ppm/K
1) according to EN 60746-1, at nominal operating conditions
2) ± 1 count
3) plus sensor error