Portavo® 904 OXY

Portable Meter

User Manual

www.knick.de

Read before installation.  

Keep for future use.

Содержание Portavo 904 OXY

Страница 1: ...Portavo 904 OXY Portable Meter User Manual www knick de Read before installation Keep for future use ...

Страница 2: ......

Страница 3: ... shipment The consignment must always be accompanied by a corresponding return form to prevent service employees being exposed to potential hazards Further information can be found at www knick de Disposal The local codes and regulations must be observed when disposing of the product Repair The meter cannot be repaired by users For inquiries regarding repairs please contact Knick Elektronische Mes...

Страница 4: ... a Sensor 13 Switching On the Meter 14 Icons 14 Configuration 15 Calibration 16 Measuring 21 Data Logger 22 Operating Modes of the Data Logger Logger Type 23 Data Logger Menu 25 Configuring the Data Logger 25 Configuring the Logger Type 26 Starting the Data Logger using CONT 27 Starting the Data Logger using START 27 Displaying the Logger Data 28 Stopping the Data Logger 29 Clearing the Data Logge...

Страница 5: ...mperature Calibration 31 Enabling Options TAN Input 32 Access Codes for CONF CAL and Data Logger 33 Inputting the Rescue TAN 34 Paraly SW 112 PC Software 35 Error and Status Messages 36 Sensoface Messages 37 Error Messages 38 Product Line 39 Sensors 39 Accessories Options 39 Specifications 41 Index 43 ...

Страница 6: ...er 4 batteries AA Carrying strap USB cable 1 5 m Quickstart overview for attaching to the inside of the protective cover German English French Safety guide Quickstart instructions in various languages Test report 2 2 according to EN 10204 User manuals the Paraly SW 112 PC software and other product information can be downloaded from www knick de ...

Страница 7: ...esistance and dimen sional stability even when exposed to extreme moisture Overview of the Portavo 904 OXY Quiver Scratch proof clear glass display perfectly readable even after years Very long operating time with one set of batteries 4x AA or use of a Li ion battery for reliable operation even at high or very low operating temperatures Data logger with 5000 values Micro USB port for communication...

Страница 8: ...more Memosens allows the storage of calibration data operating times and other data which is available and can still be used when the sensor is connected to another Memosens capable device Sensoface Sensoface provides quick information on the sen sor condition The three smiley faces as shown on the right represent the sensor condition during measurement and after a calibration When the condition d...

Страница 9: ...f the cover explains the control functions and device messages Hook A fold out hook on the back allows the meter to be suspended This leaves your hands free for the actual measurement The nameplate is located beneath the hook Protective Cover and Hook Combined The two parts can be combined to form a benchtop stand enabling convenient and fatigue free work with the device at a laboratory table or d...

Страница 10: ...information is provided by means of icons Sensoface battery icon etc Some typical displays are shown here Measuring display of measured value tempera ture date and time Calibration step 1 calibration method in air Logger data display of measured value memory location temperature date and time Clock display of hours and minutes seconds and date Calibration step 2 adjusting the relative humidity ...

Страница 11: ...a see Start up meas Switches the meter on Activates measuring mode Data logger stopping cal Start calibration set Activates configuration Confirms entries clock Displays time and date allows setting the clock using set RCL View stored values STO Holds and saves a measured value allows setting and starting of the logger by pressing set page 22 When this icon is displayed you can use the arrow keys ...

Страница 12: ...ntended function prolonged storage at temperatures above 70 C 158 F after severe transport stresses In this case a professional routine test must be performed This test should be carried out at our factory A special lithium ion battery ZU 0925 suited to the battery compartment is avail able for the Portavo 904 Only this battery type can be charged directly from the USB port A battery icon in the d...

Страница 13: ...power on a separate temperature probe is automatically recognized If you want to replace the temperature probe you must switch off the meter and then switch it on again Commissioning Memosens sensors have a cable coupling which allows for replacement of sensors while the cable remains connected to the meter The connecting cable is connected to socket b M8 4 pins or e M12 8 pins a b c d e Connectio...

Страница 14: ...justment data before it switches to measuring mode Depending on the connected sensor and the specific measuring task several steps for configuration and calibration must be performed as described on the following pages Battery charging level Temperature detector Icons Important information about the state of the device Process variable Commissioning Measured value Date Logger active Time Logger ty...

Страница 15: ...row keys confirm by pressing set LIQU GAS Sat Air Concentration in mg l if MEAS MEDIUM LIQU OFF date time date time 0 0 100 0 if MEAS MEDIUM GAS 0 4000 m 0 0 45 0 g kg AIR CAL ZERO CAL DATA INPUT FREE CAL OFF 1 99 days OFF 12h 6h 1h 0 1h C F 24h 12h DD MM YY MM DD YY Enter TAN to enable option see page 32 OFF 0000 0001 9999 with option 001 SOP only see page 33 NO YES reset to factory settings Note...

Страница 16: ...idity value cal xxx nA blinks Hourglass blinks Calibration is performed Please wait CAL DATA Date SLOPE xxx nA Slope calibration data is displayed The meter then automatically returns to measuring mode AIR CAL Calibration Calibrating the slope in air The calibration method is selected in the configuration menu Calibration Note To abort calibration you can press meas at any time ...

Страница 17: ...on is performed Please wait Date ZERO P xxx nA Zero calibration data is displayed Date SLOPE xxx nA Slope calibration data is displayed The meter then automatically returns to measuring mode ZERO CAL Calibration Zero calibration with oxygen free medium The calibration method is selected in the configuration menu Note To abort calibration you can press meas at any time Calibration ...

Страница 18: ...Measurement cal CAL DATA INPUT xx blinks nA ZERO POINT Use to set the known value for the sensor zero point cal xxx blinks nA SLOPE Use to set the known value for the sensor slope cal Calibration is performed The meter then automatically returns to measuring mode Note To abort calibration you can press meas at any time Calibration ...

Страница 19: ...nt temperature measured by sensor offset display in K cal Temperature setpoint value blinks Use to set the temperature setpoint value cal Calibration is performed the offset value is indicated The meter then automatically returns to measuring mode TEMP OFFSET OFFSET option Temperature calibration offset Selected in the configuration menu Calibration Note To abort calibration you can press meas at ...

Страница 20: ...tion menu Measurement cal CAL AIR CAL blinks Use to set the required calibration method AIR CAL ZERO CAL DATA INPUT cal Perform the selected calibration as described on the previous pages The meter then automatically returns to measuring mode Calibration Note To abort calibration you can press meas at any time ...

Страница 21: ...ons for the sensor 2 Switch the meter on using the on off or meas key 3 Depending on the measurement method and the sen sor used immerse the sensing part of the sensor in the medium to be measured 4 Watch the display and wait for the reading to stabilize 5 By pressing the STO key you can hold and save a measured value see data logger page 22 Measurement can also be controlled via the Paraly SW 112...

Страница 22: ...ta are recorded primary value temperature time stamp and device status Option 001 SOP can be used to set up an access lock for the data logger which in the absence of an access code allows only logger data to be displayed see page 31 The Paraly SW 112 PC software allows convenient management of the data logger It is always the currently selected process variable which is recorded The STO icon and ...

Страница 23: ...he last recorded value 1 Manual Logging when Logger is Deactivated Measurement Logger deactivated STO Measured value is maintained Proposed address blinks address of the last recorded value 1 If desired Select a start address using STO Measured value is saved to the desired address e g for overwriting an incorrect measurement Example Logger type INT Data Logger Interval INT In this mode the measur...

Страница 24: ...Delta Delta Delta Difference Interval Combined DIFF INT When the delta range related to the last DIFF entry is exceeded a new entry is creat ed example entry A and the delta range is displaced upwards or downwards by the delta value As long as the measured value remains within the delta range logging is performed at the preset interval The first DIFF entry is automatically created when the data lo...

Страница 25: ... Data Logger Configuring the Data Logger Prerequisite The data logger is stopped press meas Measurement STO Measured value is maintained set Logger CONT blinks Logger START blinks Logger DEL blinks Logger SET blinks set Logger Current logger type blinks Select desired logger type using DIFF INT DIFF INT or SHOT set Select the appropriate parameters using and confirm each selection by press ing set...

Страница 26: ... 00 mg l GAS Delta OFF 0 001 9 999 1 000 Delta C F OFF 0 1 50 0 C 1 0 C OFF 0 1 100 0 F 1 0 F INT Interval h mm ss 0 00 01 9 59 59 0 02 00 DIFF INT DIFF See logger type DIFF INT See logger type INT SHOT Currently selected process variable is recorded 1 Process variables dependent on configuration see page 15 Configuring the Logger Type ...

Страница 27: ... to the selected start address exception SHOT FREE MEMORY is displayed LOGGER and active logger type icons are displayed Starting the Data Logger using START Prerequisite Data logger is configured All existing entries are deleted The start address for saving the values is 0001 Every time the meter has been switched off the data logger must be restarted exception SHOT Measurement STO Measured value...

Страница 28: ...onvenient management of the data logger Measurement RCL The RCL icon and the last record ed value is displayed Use to select the desired address Empty memory locations will also be displayed RCL or meas Return to measurement Example Measured value stored at location 0026 Example Empty memory location 0004 Data Logger ...

Страница 29: ... SET set All stored data are deleted 0000 DELETED is displayed Stopping the Data Logger You can stop the data logger at any time by pressing the meas key Measurement logger activated meas Data logger is stopped LOGGER and active logger type icons are no longer displayed It is still possible to hold a measured value by pressing STO and send it to any desired address Data Logger ...

Страница 30: ...lock proceed as follows Display of time date Hour display blinks SET HOUR Set value Minute display blinks SET MINUTE Set value Second display blinks and shows 00 Clock is started the seconds count up Year display blinks SET YEAR Set value Month display blinks SET MONTH Set value Day display blinks SET DAY Set value Display of corrected time date ...

Страница 31: ...are see page 33 Configuration SETUP CODE Calibration CAL CODE Data logger LOGGER CODE Without entry of an access code the data logger will only display logger data RCL Temperature Calibration also separately available as Option 002 TEMP CAL Option 002 TEMP CAL Temperature Calibration For Memosens sensors you can perform a 1 point calibration of the internal temperature detector See page 19 for a d...

Страница 32: ...eys to select the TAN TEMP CAL function for exam ple where you can enter the TAN for enabling the option set TAN TEMP CAL Press the set key set Enter the TAN code First digit blinks Set value set Next digit blinks Set value set Set value press set to save the TAN After correct input of the TAN the device signals PASS The option is now available Enabling Options TAN Input Options ...

Страница 33: ...inks Set value set Set value press set to save the configuration access code When accessing the configuration menu you will be prompted to enter an access code If you want to set a code for access to calibration or the data logger select CAL CODE or LOGGER CODE and proceed as described above Note Functions are accessible to anyone with access code 0000 Important Note If you lose the SETUP access c...

Страница 34: ...SCUE For this purpose please have the serial number of the corresponding device to hand If you have any questions please contact Knick Elektronische Messgeräte GmbH Co KG using the contact details provided on the last page of this document The menu for input of the rescue TAN appears if the SETUP access code is incorrectly entered three times ...

Страница 35: ...ive Windows user interface Easy configuration and management of several meters Display of device and sensor information Convenient management and evaluation of the data logger Export function for Microsoft Excel Print function Upgrade downgrade of device firmware Note Prior to upgrading downgrading the device firmware Portavo is reset to its factory settings Make the following backups prior to upg...

Страница 36: ... Measurement can still be performed After a calibration the corre sponding Sensoface icon friendly neutral sad is shown together with the calibration data Otherwise Sensoface is only visible in measuring mode The most important error messages and Sensoface info messages are shown on the inside of the protective cover A com plete list of messages and their meanings is provided in the following tabl...

Страница 37: ...nsoface Meaning Sensor is okay Calibrate the sensor soon Calibrate or replace the sensor The neutral and sad Sensoface icons are accompanied by an INFO message to give a hint to the cause of deterioration Sensoface Message Cause INFO 1 Calibration timer INFO 5 Zero Slope INFO 6 Response time INFO 8 Leakage current ...

Страница 38: ... not help replace the sensor ERROR 5 Slope too high low ERROR 11 Measured value unstable Stability criterion not met Leave the sensor in the liquid until the temperature is stable If this does not help replace the sensor ERROR 14 Time and date invalid Set time and date ERROR 18 Configuration invalid Restart reset to factory settings Setup DEFAULT YES configure and calibrate If this does not help s...

Страница 39: ...U6953 Maintenance kit for SE715 1 MS electrolyte 3 membrane caps ZU0879 Flow through cell for SE715 1 MS oxygen sensor ZU1014 O2 electrolyte ZU0565 Measuring cable with M8 connector for sensors with Memosens connector Length 1 5 m 4 92 ft CA MS 001XFA L Length 2 9 m 9 51 ft CA MS 003XFA L Measuring cable with M12 connector for sensors with Memosens connector Length 1 5 m 4 92 ft CA MS 001XDA L Len...

Страница 40: ...erification temperature detector adjustment in the Memosens sensor offset correction SW P001 Temperature detector adjustment in the Memosens sensor offset correction SW P002 Paraly SW112 PC software for configuration and firmware updates Free download from www knick de Product Line ...

Страница 41: ...iendly neutral sad Status indicators For battery condition logger Notices Hourglass Keypad on off cal meas set STO RCL clock Data logger With up to 5000 memory locations Recording Manual interval or event controlled Communication USB 2 0 Profile HID driverless installation Usage Data exchange and configuration via Paraly SW 112 PC software Diagnostic functions Sensor data Manufacturer sensor type ...

Страница 42: ... chargeable Operating time Approx 500 h alkaline Rated operating conditions Ambient temperature 10 C 55 C 14 122 F Transport Storage temperature 25 70 C 13 158 F Relative humidity 0 95 short term condensing allowed Housing Material PA12 GF30 silver gray RAL 7001 TPE black Protection IP66 67 with pressure compensation Dimensions Approx 132 x 156 x 30 mm Weight Approx 500 g ...

Страница 43: ...icon 12 Battery replacement 12 Benchtop stand 9 C CAL CODE 31 Calibration access control 31 cal key 11 Charge level of batteries 12 Clearing the data logger 29 Clock 30 clock key 11 Commissioning 12 Configuration access control 31 Configure data logger 25 Connecting a sensor 13 Connecting cable for Memosens 13 Connections 13 Connection USB battery 12 Continuous recording of measured values 23 CONT...

Страница 44: ... 24 Device configuration 15 Device messages 36 Device properties 7 Difference Interval data logger mode 24 Difference data logger mode 24 Display 10 Display icons 14 Displaying the data logger 22 Displaying the time and date 30 Disposal 3 E Electrolyte accessory 39 ERROR error codes 38 Error messages 36 Error messages overview 38 F Features 7 Field case accessory 39 FREE CAL calibration 20 FREE CA...

Страница 45: ...um ion battery 12 Logger 22 LOGGER CODE 31 Logger type configuration 26 Logger type data logger modes 23 M meas key 11 meas switch on 14 Measured value recording 23 Measuring 21 Memory for measured values 22 Memosens 8 Memosens cable accessory 39 Memosens connecting cable 13 Memosens sensors 13 Menu of data logger 25 Menu structure of configuration 15 Menu structure of data logger 25 Messages 36 M...

Страница 46: ...configuration 15 Overview of error messages 38 Oxy configuration 15 Oxygen free medium 17 Oxygen sensor 39 P Package Contents 6 Paraly SW 112 PC software 35 Paraly SW 112 PC software 35 Parameter setting data logger 25 Parameter settings configuration 15 Power on 14 Product features 7 Product presentation 7 Protective Cover 9 R RCL displaying the logger data 28 RCL key 11 Real time clock 7 Recharg...

Страница 47: ...tion data 15 Setting the data logger 25 Setting the time and date 30 SETUP CODE 31 SHOT data logger mode 23 Smiley 37 SOP Option Standard Operating Procedure 31 Specifications 41 Start address data logger 23 Starting the data logger using CONT 27 Starting the data logger using START 27 START starting the data logger 27 Status messages 36 STO activating the logger 23 STO key 11 Stopping the data lo...

Страница 48: ... input 32 Temperature calibration TEMP OFFSET option 19 Temperature detectors accessory 39 TEMP OFFSET OFFSET option 19 Triangle icons 11 U USB port battery 12 V Value added features 8 Viewing recorded data 28 Viewing the logger data 28 Z ZERO CAL calibration 17 Index ...

Страница 49: ......

Страница 50: ...80191 0 Fax 49 30 80191 200 info knick de www knick de Local Contacts www knick international com Copyright 2021 Subject to change Version 2 This document was last updated on March 31 2021 The latest documents are available for download on our website under the corresponding product description TA 209 4OX KNEN02 ...