30
Specifications
Permissible process pressure and
temperature during movement
10 bar (up to 140 °C)
Permissible process pressure and
temperature, statically in SERVICE position
16 bar (at 0 ... 40 °C)
Ambient temperature
-10 ... 70 °C
Ingress protection
IP 66
Permissible pressure for probe control
(4) 5 ... 7 bar
*
Quality of compressed air
Standard
according to ISO 8573-1:2001
Quality class
3.3.3 or 3.4.3
Solid contaminants
3 (max.
5
µm, max. 5 mg/m3)
Water content for temperatures > 15 °C
Class 4, pressure dew point 3 °C or below
Water content for temperatures 5 ... 15 °C
Class 3, pressure dew point -20 °C or below
Oil content
Class 3 (max. 1 mg/m3)
Sensors
with solid electrolyte
Ø 12 mm, length 425 mm with temp detector,
PG 13.5 thread
with liquid electrolyte
Ø 12 mm, length 450 mm with temp detector
Optical sensors
Ø 12 mm, Ø 12.7 mm (1/2“), special version B required
Process adaptations
See product coding
Connections
Outlet
Socket for Unical media hose
for pressurized sensors
Hose connection NW 6
Pressure in sensor chamber 0.5 … 1 bar above process pressure
(max. 7 bar)
for compressed air,
rinse and calibration media
(control air for retractable fitting)
for Unical multiplug
Immersion depths / Dimensions
See dimension drawings
Process-wetted materials
See product coding
Control pressure (bar) Process pressure (bar)
5
7
6
8
7
10
* Control pressure required for process pressure
Содержание Ceramat WA 154
Страница 1: ...User Manual English Latest Product Information www knick de Ceramat WA 154 Sensor Lock Gate ...
Страница 27: ...27 Installation Dimensions Approx 600 mm clearance for sensor installation Solid Electrolyte ...
Страница 28: ...28 Installation Dimensions Approx 800 mm clearance for sensor installation Liquid Electrolyte ...
Страница 42: ......