13
Information on Installation
For operation of the Ceramat, you must connect control air, rinsing or calibration media, and the
electrical˜check-back signal for indicating the probe position.
There are two possible ways to do so.
When the Ceramat is operated with a Unical or Uniclean electro-pneumatic controller and the Protos measur-
ing system, the cables and tubings for air pressure, rinsing/calibration media and check-back are combined in
a single hose with just one plug connection (multiplug). This hose is referred to as media connection.
This media connection is installed on the Ceramat together with the outlet hose.
When you do not use a probe controller (Unical or Uniclean and Protos measuring system), you can connect
the supply lines for control air, rinsing/calibration media and electrical check-back to the Ceramat with a free
hose connection via the ZU 0631 standard-media interface.
Connection for operation with
Unical/Uniclean und Protos
Connection for operation with
ZU 0631 standard media interface