45
Configuring the Current Output
HOLD active
21
4
HOLD active
Output current
[mA]
Output signal for HOLD
FIX setting = 21.0 mA
Output signal for HOLD
LAST setting
Output signal during HOLD:
Menu item
Action
Choices
Output current during
error message
The output current can be set to
22 mA in the case of error mes-
sages or error messages.
Select ON or OFF using
keys.
Confirm by pressing enter
OFF / ON
Output current during
Sensoface messages
The output current can be set to
22 mA in the case of Sensoface
messages.
Select ON or OFF using
keys.
Confirm by pressing enter
OFF / ON
Output current during
HOLD
LAST: During HOLD the last
measured value is maintained at
the output.
FIX: During HOLD a value (to be
entered) is maintained at the
output.
Select using
Press enter to confirm.
LAST/FIX
Output current for HOLD
FIX
Only with FIX selected:
Enter current which is to flow at
the output during HOLD
Enter value using
keys.
Press enter to confirm.
00.00...22.00 mA
(21.00 mA)