Safety 

 

Vacuum Pump System SC 920G 

Translation of original Operating Instructions, English, KNF 121308-121311 07/18 

3.  Safety 

The vacuum pump system is built according to the generally recog-
nized rules of technology and in accordance with the occupational 
safety and accident prevention regulations. Nevertheless, dangers 
can result during their use which lead to injuries to the user or 
others, or to damage to the vacuum pump system or other property. 

Only use the vacuum pump system when it is in a good technical 
and proper working order, in accordance with its intended use, 
observing the safety advice within the operating instructions, at all 
times. 

Make sure that only trained and instructed personnel or specially 
trained personnel work on the vacuum pump system. This 
especially applies to assembly, connection and servicing work. 

Make sure that the personnel has read and understood the opera-
ting instructions, and in particular the "Safety" chapter. 

Observe the accident prevention and safety regulations when per-
forming any work on the vacuum pump system and during opera-
tion. 

Do not expose any part of your body to the vacuum. 

Open housing parts with notice sticker (see fig. 1) only after sepa-
rating mains plug from power source. 

 

 

Ensure that the employees check that the hand terminal is the right 
one for this particular system before using a SC 920G vacuum 
pump system. Vacuum pump systems are equipped with a paging 
system for this purpose (see Actuating the vacuum pump system, 
page 29).

 

When transferring dangerous media, observe the safety regula-
tions when handling these media. 

Be aware that the vacuum pump system is not designed to be 
explosion-proof. 

Make sure the temperature of the medium is always sufficiently 
below the ignition temperature of the medium, to avoid ignition or 
explosion. This also applies for unusual operational situations. 

Note that the temperature of the medium increases when the pump 
compresses the medium. 

Hence, make sure the temperature of the medium is sufficiently 
below the ignition temperature of the medium, even when it is 
compressed to the maximum permissible operating pressure of the 
vacuum pump system. The maximum permissible operating pres-
sure of the vacuum pump system is stated in the technical speci-
fications (see chapter 4, page 8). 

If necessary, consider any external sources of energy, such as 
radiation, that may add heat to the medium. 

In case of doubt, consult the KNF customer service. 

Personnel 

Working in a safety- 

conscious manner 

 

Fig. 1: Notice sticker 

Correct match between hand 

terminal and vacuum pump 

system 

Handling dangerous media 

Handling flammable media 

Содержание SC 920G

Страница 1: ...ons english Operating Instructions Read and observe this Operating Instructions Vacuum pump system with wireless remote control SC 920G KNF Neuberger GmbH Alter Weg 3 D 79112 Freiburg Germany Phone 49...

Страница 2: ...o PC Power supply for hand terminal Operating Instructions Abbreviated instructions CD with digital Operating Instructions and software for operating the vacuum pump system with a PC Allen key for tra...

Страница 3: ...6 1 Installation 17 6 2 Connection 18 7 Operation 19 7 1 Start up 19 7 2 Taking system out of operation 22 8 Operating vacuum pump system 23 8 1 Hand terminal 23 8 1 1 General functions and displays 2...

Страница 4: ...ailure to do so may result in damage or injury 1 2 Symbols and Markings Warning WARNING A danger warning is located here Possible consequences of a failure to observe the warning are specified here Th...

Страница 5: ...he pump The high performance condenser must be installed on the outlet side of the pump if it is installed on the inlet side there is a danger of implosion Observe the correct usage of the gas and coo...

Страница 6: ...ains plug from power source Ensure that the employees check that the hand terminal is the right one for this particular system before using a SC 920G vacuum pump system Vacuum pump systems are equippe...

Страница 7: ...The vacuum pump system SC 920G conforms to the Directive 2011 65 EU RoHS2 The vacuum pump system SC 920G conforms to the safety regulations of the EC Directive 2014 30 EU concerning Electromagnetic Co...

Страница 8: ...open gas ballast Delivery rate at atm pressure l min 0 21 controllable Max permissible pressure at inert gas connection bar g 0 1 Ambient and media temperature Permissible ambient temperature 10 C to...

Страница 9: ...rature 15 C to 20 C Coolant connections on high peformance condenser For tube ID 8 mm Coolant wetted surface cm2 Min 460 Electrical data vacuum pump system Voltage V 100 240 10 Frequency Hz 50 60 Max...

Страница 10: ...ding on number of entries and data transmission Charging time about 7 h Tab 3 Applies to standard included batteries When charging the hand terminals of the SC 920G vacuum pump system use only the ori...

Страница 11: ...condenser 5 Gas ballast switch 6 Ventilation connection 7 Vacuum chamber connection gas inlet 8 Hexagonal gland 9 Pump 10 Base 11 Flask clamp for 12 12 Suction side condensation collection flask coat...

Страница 12: ...f the pump suctioned from the vacuum chamber The collection flask is coated implosion protection and fastened to the vacuum pump system via a flask clip The high performance condenser at the pump outl...

Страница 13: ...quickly as possible Option Automatic with time after the start of the process Automatic detection of the boiling pressure If the set limit pressure is reached without a vapor pressure the program goe...

Страница 14: ...cess control active control of the distillation using the setpoint pressure In order to change to another operating mode the process is first stopped and then restarted in the new operating mode The g...

Страница 15: ...from the holder see chapter 8 1 1 page 23 on the vacuum pump system This provides a convenient way to operate the vacuum pump system when it is in a cabinet or under a closed fume hood Whenever the h...

Страница 16: ...upward stroke the diaphragm presses the medium out of the pump head via the outlet valve 1 The transfer chamber 3 is hermetically separated from the pump drive 7 by the diaphragm 5 5 Gas ballast WARN...

Страница 17: ...he vacuum pump system Protect the vacuum pump system from dust Protect the vacuum pump system from vibrations and jolts Only connect components to the vacuum pump system which are designed for the pne...

Страница 18: ...ntity of gas at the gas outlet and do not install any compo nents that hinder the gas flow Safely discharge gas exhaust so that no gas can escape into the ambient air Make sure that the high performan...

Страница 19: ...er is not pressure proof Vacuum pump system is at room temperature Vacuum pump system and hand terminal belong together No reactive explosive or otherwise hazardous mixtures may be produced when venti...

Страница 20: ...to indicate which hand terminal belongs to which vacuum pump system see chapter 11 Ordering information WARNING Uncontrolled operation may result in personal injury and damage to the vacuum pump syste...

Страница 21: ...e condenser s gas outlet is not blocked In order for the high performance condenser to recover solvent from the delivered gas it must be cooled by means of a cold water supply or recirculating cooler...

Страница 22: ...ing aggressive media flush the vacuum pump system prior to switch off to increase the service life of the diaphragms see chapter 9 2 1 page 33 Switch off vacuum pump system with mains switch see fig F...

Страница 23: ...b a touchscreen The rotary push knob has the following functions Rotate Changes pump capacity or the selected pressure setting depending on selected operating mode Press Interrupts the active process...

Страница 24: ...e hand terminal is switched on or off Every time you use a hand terminal check to ensure a correct match between the terminal and the vacuum pump system Use the paging system for this purpose see Actu...

Страница 25: ...essure is applied as the setpoint pressure Fig 9 Menus and buttons on the touchscreen 8 1 2 Operation Menu language When the start page appears after switching the hand terminal on choose between the...

Страница 26: ...ging operating mode see Fig 11 at the top Operating mode can be changed only when no process is active Switching to manual process control Break Press briefly on the rotary push knob Manual process co...

Страница 27: ...ventilation valve remains open If the ventilation valve is permanently open OFF button shown in display it can be closed again with the OFF button Opening and closing coolant valve accessory Pressing...

Страница 28: ...e is applied as the new setpoint pressure and the vacuum pump system regulates to this pressure The current setpoint pressure is shown in the display in place of the arrow If you press again on the se...

Страница 29: ...t a selected line To change columns e g from t to p mbar Press the rotary push knob To change an entry Turn the rotary push knob After the last column the display automatically returns to line selecti...

Страница 30: ...r the active process Fig 8 5 page 25 in the hand terminal display the LED next to the page button Fig 15 page 32 on the vacuum pump system will blink The other way around the hand terminal will answer...

Страница 31: ...partly or total defect of the hand terminal Only handle the hand terminal at an ESD protected area EPA by a qualified person under the IEC 61340 5 1 directive 1 Loosen the four screws on the underside...

Страница 32: ...close the ventilation valve Open and close the high performance condenser s coolant valve accessory Call the hand terminal paging the hand terminal will answer with a signal tone 1 Stop process button...

Страница 33: ...or explosive mixtures can form Before switching off the vacuum pump system separate it from the vacuum chamber and flush it with air if necessary for safety rea sons with an inert gas under atmospher...

Страница 34: ...h KNF 121308 121311 07 18 Pressure side 1 Grasp the collection flask Fig 2 15 page 11 and simulta neously remove attachment clamp 16 pull out collection flask 2 Dispose of contents in collection flask...

Страница 35: ...and pump Depending on the substance transferred caustic burns or poisoning are possible Wear protective clothing if necessary e g protective gloves Flush vacuum pump system before replacing the diaphr...

Страница 36: ...rod 21 do not fall into the pump housing If diaphragm spacers should adhere to the diaphragms take them off and put them on the thread of the corresponding connecting rod In order to ensure the pneuma...

Страница 37: ...ion see fig 17 7 Dispose of the old diaphragms and valve plates sealings properly Mount pump head 1 Place head plate 9 on intermediate plates 10 13 and 14 in the position indicated by the drawing line...

Страница 38: ...uum Pump System SC 920G 38 Translation of original Operating Instructions English KNF 121308 121311 07 18 Final steps 1 Reconnect vacuum pump system tubing 2 Reconnect vacuum pump system to the electr...

Страница 39: ...substance transferred caustic burns or poisoning are possible Wear protective clothing if necessary e g protective gloves Flush vacuum pump system before replacing overpressure valve see chapter 9 2 1...

Страница 40: ...fuse s of the vacuum pump system Push the locking clip on the fuse drawer down Fig 3 4 p 12 and pull the drawer out Replace the defective fuse s for fuse specifications see chapter 4 1 page 8 for fuse...

Страница 41: ...ystem up in evacuation mode 100 capacity Observe the safety instructions in chapter 3 2 Using air Open gas line Fig 2 7 p 12 to the surrounding environment and start system up in evacuation mode 100 c...

Страница 42: ...is sitting correctly replace if necessary For ordering number see chapter 11 page 45 Tab 7 Hand terminal Problem Cause Solution Hand terminal cannot be removed from vacuum pump system Transport fasten...

Страница 43: ...is currently in operation Use the page function see Actuating the vaccum pump system p 29 to check whether the right hand termi nal is being used Acoustic alarm sounds on hand terminal Batteries low C...

Страница 44: ...as been changed Set desired unit of pressure Leaks in system See Tab 8 Required vacuum not generated even though pump is running Recalibration of pressure sensor required Contact KNF Service Tab 9 Fau...

Страница 45: ...ndenser 047807 Battery set for hand terminal see chapter 8 1 3 page 31 117427 Power supply for hand terminal 312797 USB cable 136174 Mains cable D 026363 Mains cable CH 027523 Mains cable GB 029866 Ma...

Страница 46: ...nt valve 117121 Color sticker for indicating correct match between hand terminal and vacuum pump system 117433 Chemical resistant protective film for hand terminal display 117407 Wall fixture hand ter...

Страница 47: ...tamination form For optimal processing of a return a copy of this declaration should be sent in advance via e mail regular mail or fax to KNF Customer Service refer to final page for address In order...

Страница 48: ...Health and safety clearance and decontamination form Vacuum Pump System SC 920G KNF worldwide Find your local KNF partner on www knf com 13 Health and safety clearance and decontamination form...

Отзывы: