8. Servicing
8.1. Servicing schedule
Component
Servicing interval
Pump
-
Regular inspection for external
damage or leaks
Pump head
-
Clean if the flow rate decreases, the
pump does not work or no vacuum
is created (Section 8.2)
Diaphragm, valve
plates and seals
-
Change as soon as pumping
capacity decreases, preferably
sooner
Tab. 13
8.2. Cleaning
WARNING
Health hazard due to dangerous substances
in the pump!
Depending on the substance transferred, risk of
caustic burns or poisoning.
➔
Wear protective clothing if necessary, e.g.
protective gloves.
➔
Rinse the pump with a neutral liquid and pump
empty.
8.2.1. Flushing the pump
➔
If pumping aggressive media, KNF recommends flushing the
pump with air under atmospheric conditions for several
minutes prior to switch off (if necessary for safety reasons: use
an inert gas). This will extend the service life of the diaphragm.
8.2.2. Cleaning the pump
➔
Where possible, wipe the components with a soft dry cloth. Do
no use cleaning solvents as these may corrode plastic parts.
➔
If there is compressed air available, blow off the separate
parts.
▪
Pump must be switched off and mains plug removed from
the socket.
▪
The pump must be free of any hazardous substances.
▪
Hoses must be disconnected from the pump head.
▪
We recommend replacing the diaphragm when the head
parts are removed.
Qty.
Tool
1
Torxplus 10 IP screwdriver
Tab. 14
Information on procedure
Prior requirements
Tools