7.4 Connecting water cooling (optional)
) can increase the service life
of the diaphragm, particularly with high pressures or high ambient
temperature.
Operating parameters
Parameter
Value
Water temperature [°C]
+ 5 to + 30
Water pressure [bar gauge]
< 1.0
Water flow rate [l/min]
> 1.0
Tab.18 Recommended parameters for connecting the water cooling
1. Connect the water cooling system to the provided hose connections.
à
Safely drain the water discharge.
The flow direction is allowed in both directions.
Mounting dimensions For mounting dimensions, see the following dimensional drawings:
Fig.6 Mounting dimensions N680.1.2 EX