Installation
Diaphragm Vacuum Pumps N 012, N 024, N 036
6 Translation of original Operating and Installation Instructions, english, KNF 121222-121485 08/11
3. Connect the mains cables to the connections L 1 and N of
the luster terminal (see figs. 3a, 3b or 4).
4. Connect the earth (ground) wire to the pump motor.
5. Place the electronics housing or the terminal cover back in
position and screw it tight; make sure the housing sealing is
properly sited.
6. Only .26 versions (pumps with heated heads; electronically
controlled): Connect the serial interface cable to an avail-
able, active COM port on the PC (to operate the pump with-
out PC software, the interface cable can be detached on
terminals X3/1-5 and removed; use the protective cap sup-
plied to cover the unused cable fitting).
Fig. 3b: Electronical connection of N 036.0 _T. 11 E
Fig. 3a: Electrical connection
(.11-designs: with themostatically
regulated head heating) except
for N 036_T.11 E)
Содержание N 012 AT. 16 E