Initial start-up
Installation Instructions FP 70
KNF Flodos 330352_MA_FP70_EN_00
10
4 Initial start-up
The product must not be started up until it has been ensured that
the machine in which the product is to be installed meets the re-
quirements of the Machinery Directive 2006/42/EC, if applicable.
WARNING
Danger that the fluid system could rupture due to excess
pressure
The pump builds up pressure. With a closed system, the max.
permitted operating pressure can be exceeded. This may cause
injuries to personnel, or damage to the pump or the system.
➢
Prevent operation against a closed system.
➢
In the case of parts that are in contact with the fluid, use
only those that are designed to withstand at least the oper-
ating pressure of the pump.
➢
If necessary, take suitable measures to limit the maximum
system pressure.
CAUTION
Danger of unexpected chemical reactions with water
Residues of water in the pump that come from testing in the fac-
tory may react with the transferred medium.
➢
Before starting the pump for the first time, rinse it with a me-
dium that is uncritical as regards water.
VORSICHT
Danger of automatic restart
Pumps with brushless DC-motor have a thermal overload protec-
tion with automatic recovery to protect the motor from overheat-
ing. After cooling down the motor starts automatically.
➢
Ensure that this does not cause a hazardous situation