Operating the Valve Drive S 6/-12/-16
Operating the Valve Drive S 6/-12/-16
Power supply
The valve drive is equipped with a universal power supply unit which operates in a
range of 90 to 260 Volts AC. A manual setting of the supply voltage is, therefore,
not required.
Make sure that the main power supply is properly grounded and the correct
power cable is used.
The power supply for the drive is not included in the housing. Please connect the
power supply to the drive (item
4
in Abb. 4). Then connect the power cable to the
power supply and the socket.
If required connect the ground connector
3
to the corresponding ground contact of
any other
grounded
instrument.
Wait for at least 30 seconds before restarting the drive after switching it off.
Otherwise an error message will be displayed.
Mounting the valve
The Smartline Electric Valve Drives S 6/-12/-16 have to be used in combination
with one of the KNAUER-valves. For instructions on how to operate the valves
refer to the corresponding valve manuals.
SOP 1
Mounting the valve to the electric valve drive
1. Fix the adapter plate with the two M4 screws at the rear side of the valve.
2. Set the drive on position 1 respectively LOAD.
3. Push the valve on the driving axle on the front side of the instrument.
4. Take care that the mark on the valve matches this one above the wheel. The
valve position 1 must be on top.
5. Fix the valve and adapter plate with the two mounting screws.
If the position was not set on LOAD or 1, you will get the error message E 4
when starting the drive.
Configuring the Smartline Electric Valve Drive
At the bottom of the instrument the DIP-switches 1-4 (Abb. 5) are arranged. They
are needed to adapt the drive to the actual mounted valve.
1 2 3 4
ON
Fig. 5 DIP-switches 1-4 at the drive bottom
DIP 1 adapts the two position or multi-position valve (S 6).
The valve drive S 12, exclusively used to control the 13-port/1-channel
switching valve, can only be operated with the DIP 1 switch in the ON
position!