General Description
5
General Description
The Smartline Electric Valve Drives S 6, S 12, and S 16 are in design and
operation for the most part identical. They are devised for automatic control of the
6-port/3-channel-injection valve, 6-port/2-channel-switching valve, 7-port/1-
channel-switching valve by S 6 drive, the 13-port/1-channel-switching valve by
S 12 drive, and the 17-port/1-channel-switching valve by S 16 drive respectively.
The valve drive has to be adapted to the used valve by means of a set of DIP
switches at the bottom of the drive housing.
All KNAUER valve drives can be integrated into the Smartline Assistant 6000.
Valves with 2 positions
The 6-port/3-channel-injection valve and the 6-port/2-channel-switching valve can
be switched between the LOAD and INJECT position. Switching the positions can
be done with a 60° turn forwards or backwards.
Valves with 6 positions
The 7-port/1-channel-switching valve can be switched to 6 positions. Each switch
command will result in a 60° turn, so that the 6
th
switch command returns to the
first position.
Valves with 12 positions
The 13-port/1-channel-switching valve can be switched to 12 positions. Each
switch command will result in a 30° turn, so that the 12
th
switch command returns
to the first position.
Valves with 16 positions
The 17-port/1-channel-switching valve can be switched to 16 positions. Each
switch command will result in a 22.5° turn, so that the 16
th
switch command returns
to the first position.
Using remote or software control valves with more than two positions can be
switched to any one directly.
Preparing the Electric Valve Drives S 6/-12/-16 for Operation
Unpacking
After unpacking, please check the device and accessories thoroughly for any
damage that may have occurred during transportation. If necessary, contact the
carrier to claim any damages.
Use the “Standard delivery list“ to check if the electric valve drive is complete.
Please contact our service department if anything is missing or if you need support.
It is important to fill out the guarantee card and return it to us as soon as possible.
Please remove the protection foil from the display!
Standard Delivery List
1
Smartline Electric Valve Drives S 6, S 12, or S 16
1
Power supply unit 24V/50W
1
Power supply cable, 230V
1
Adapting plate 70x50 Al
1 Set
Tools
1
RS-232 cable
1
WAGO Terminal Strip
1,5 m
Set signal conductors
1
User Manual
Содержание Smartline S12
Страница 2: ......