16
Troubleshooting
Troubleshooting
Error messages will be displayed in the form of „En“ (n = error code). A question
mark “?” will appear if an invalid command is entered.
Table 9 List of error codes, causes, and solutions
code
probable cause
solution
E0
valve has not switched
exchange valve sealing
E1
motor power consumption to
high
exchange valve sealing
E2
position changing to slow
exchange valve sealing
E3
DIP 3 and 4 in wrong setting
correct the settings
E4
position 1 no longer corresponds
to HOME
exchange valve sealing
DIP 1 and 2 in wrong setting
correct the settings
E5
wrong binary code
correct the settings
E6
error while reading the
EEPROM
restart the instrument
In case the purposed solutions will not solve your problems, please contact our
service department.
Spare parts and accessories
Power supply cable, 230V
M1642
RS-232 cable
A0895
Set WAGO connecting rails
A1402
Set signal conductors
A1467
Maintenance by the user
The Electric Valve Drives S 6, S 12, and S 16 do not need any maintenance. In
case of problems please contact our service department.