10
Operating the Valve Drive S 6/-12/-16
Table 4 DIP-switch positions in valve drive
S 16
ON not
allowed
DIP
1
OFF
360° valve (17-p/1-ch)
ON
binary control BIN 0-2
DIP
2
OFF
RS-232 control/ manually/
HOME/ INJ./ LOAD
DIP
3
DIP
4
ON
OFF
OUT/ BIN 3 output
DIP
3
DIP
4
OFF
ON
OUT/ BIN 3 input
The data in the above tables imply that only one or two DIP combinations are
possible for any of the valves to be controlled.
Table 5 Allowed DIP combinations
Valve
Mode
DIP position
Drive
6-port/2-channel
6-port/3-channel
analog OFF OFF ON OFF
analog ON OFF
ON
OFF
7-port/1-channel
binary ON ON ON
OFF
S 6
analog ON OFF
ON
OFF
13-port/1-channel
binary ON ON OFF
ON
S 12
analog OFF
OFF
ON
OFF
17-port/1-channel
binary OFF ON OFF
ON
S 16
DIP switch No :
1 2 3
4
6-Port-Multi Channel Valve Switch Positions :
The internal port connections of the multi-channel valves are shown in the following
table.
Table 6 Port connections in multi-channel valves
Operated Key
(INJECT)
(LOAD)
Display
I L
Connected ports
2-Channel valve
6
2
5
3
1
4
6
2
5
3
1
4
Connected ports
3-Channel valve
6
2
5
3
1
4
6
2
5
3
1
4