40
Semi-Micro Osmometer K-7400S instructions, V3700
12. Technical data
12.1 Main features
Feature
Specifications
Sample volume
50–150 µl
Solvent
Water (HPLC quality)
Measurement time
ca. 2 min
Measurement range 0–2000 mOsmol / kg
Precision
SD ≤ 4 mOsmol / kg [0–400 mOsmol / kg]
RSD ≤ 1.0 % [400–2000 mOsmol / kg]
Linearity
± 1 % [0–1500 mOsmol / kg]
± 1.5 % [0–2000 mOsmol / kg]
Calibration
Two-point calibration
(0 mOsmol / kg and one free selectable
osmolality)
optional:
Three-point calibration
(0 mOsmol /kg and 2 free selectable
osmo lalities)
Measuring
procedure
Completely automatic after sample insertion
12.2 Communication
Feature
Specifications
Port
RS-232
Control
Keypad (LED screen, 2 rows with 24 carac ters),
software