Using the EuroOsmo® 7400 Software
21
Fig. 13
Measuring window with the temperature sequence of a run
Fig. 14
Setup area of the main window
Setup
Area for entering general settings for measurement
sequences and corresponding reports.
Operator
Enter the operator’s name
Communication Port Select the COM port to which your osmometer is
connected.
Working Parameters
Monitoring
view The
entered
value
defines the scale of the monitor
by its final value in minutes. The temperature
pattern is shown continuously for the defined
period.
Measurement
view The entered value defines the scale of the
measurement window by its final value in minutes.
Freezing Level The initiation point „freeze“ can be set within the
range from -13.0 to -4.0 °C. Any decimal value will
be rounded in steps of 0.5°.
Cooling Limit The minimum temperature can be set within the
range from -18.0 to -12.0 °C. Any decimal value
will be rounded in steps of 0.5°.
Units Select the units valid for the measurement
sequence. The choice is mOsmol, °C, °F, and %. If
°F is selected the values of the freezing level and
cooling limit are automatically changed to °F.
By clicking on the Apply button the entered values
are sended to the osmometer. If any value is out of
the allowed range a corresponding message will